Operating Instructions
3-5
Press
to change view modes. For each mode, the active measurement or statistic result
is displayed on the second line of the display.
The T1-T2 view mode displays the current Channel 1 measurement minus the current
Channel 2 measurement. The display indicates T1-T2 at the left side of the display. If either
channel is not connected to a probe, or the current measurement on either channel is over-
or under-range, T1-T2 view mode is not available.
When viewing statistics, the active statistic is indicated directly below the result. The
elapsed time of the current statistics session is displayed in the lower-left corner of the
display.
Statistics are calculated continuously, beginning when the instrument is powered on or when
(1.5s)
is pressed. To pause statistics collection temporarily, press
. To resume statistics
collection, press
again.
It is important to note that changing parameter values or temperature probes will invalidate
the current statistics session. When using statistics, always begin by pressing
delete existing statistics data and initiate a new statistics session.
Press
to step through the available statistics. Statistics are displayed in the order shown
in Figure 8 below. For dual-channel models, the LCD T1 or T2 indicators are lit to identify the
channel’s statistics currently being displayed.
When using statistics, always begin by pressing
results and initiate a new statistics session.
V
IEW
M
ODE
D
ISPLAY
I
NDICATOR
D
ESCRIPTION
T1–T2
T1-T2
Current Channel 1 measurement – current
Channel 2 measurement
Minimum
MIN
Minimum temperature recorded during current
session
Maximum
MAX
Maximum temperature recorded during current
session
Average
AVG
Average of all temperatures recorded during
current session
Range
RNG
Maximum - Minimum
Standard Deviation
STDEV
Standard deviation of all temperatures recorded
during the current session
1
.
1
Standard Deviation is calculated using the population formula:
𝜎𝜎
=
�
∑
(
𝑥𝑥−µ
)
2
𝑁𝑁
Figure 7: View Modes and Statistics
Summary of Contents for HH931T
Page 39: ...Appendices B i B EXPANDED INSTRUMENT UNCERTAINTIES Thermocouple Type B ...
Page 40: ...Appendices B ii Thermocouple Type E ...
Page 41: ...Appendices B iii Thermocouple Type J ...
Page 42: ...Appendices B iv Thermocouple Type K ...
Page 43: ...Appendices B v Thermocouple Type N ...
Page 44: ...Appendices B vi Thermocouple Type R ...
Page 45: ...Appendices B vii Thermocouple Type S ...