CAUTION: Resettable fuse will not protect meter if power supply voltage
exceeds maximum voltage specified for particular model.

In order to operate flow meter within the specification the cable length for analog
0-5 Vdc output and remote LCD display may not exceed 9.5 feet (3 meters).

Use of the FMA1700A/1800A flow transducer in a manner other than that speci-
fied in this manual or in writing from OMEGA

7

, may impair the protection provid-

ed by the equipment.

2.3.1   Remote LCD Readouts

FMA1700A/1800A Mass Flow Meters are available with optional remote reading
LCD displays supplied with a three foot long wire to accommodate most applica-
tions. This configuration includes the upper block element which serves as the
LCD readout mounting.  Special lengths of remote extension wiring (up to 9.5 feet
[3 meters]) are available on request.

2.3.2  Panel Mounting Readouts

Another option for the FMA1700A/1800A Mass Flow Meter is the Panel Mounting
Remote Readout.  In this configuration the LCD readout is supplied with a three
foot long extension wire, and no aluminum housing around the LCD. The LCD
readout for panel mounting includes a bezel with two plastic screws which conve-
niently fit into a rectangular cut-out for panel mounting (see Figure 2.b). 

The power input is protected by 300 mA M (medium time-lag) resettable fuse. If a
shorting condition or polarity reversal occurs, the fuse will cut power to the flow
transducer circuit. Disconnect the power to the unit, remove the faulty condition,
and reconnect the power. The fuse will reset once the faulty condition has been
removed.

FIGURE 2.b - 

CUTOUT DIMENSIONS FOR LCD PANEL MOUNTING

Summary of Contents for FMA1700A

Page 1: ...FMA1700A FMA1800A Mass Flowmeter e mail info omega com For latest product manuals www omegamanual info User sGuide Shop online at omega com NORWALK CT TM TM...

Page 2: ...cifications without notice TM omega com info omega com Omega Engineering Inc Toll Free 1 800 826 6342 USA Canada only Customer Service 1 800 622 2378 USA Canada only Engineering Service 1 800 872 9436...

Page 3: ...eadings 5 3 Swamping Condition 6 MAINTENANCE 6 1 Introduction 6 2 Flow Path Cleaning 6 2 1 Restrictor Flow Element RFE 6 2 2 FMA1700A 1800A Series Max Flow 10 L min 7 CALIBRATION PROCEDURES 7 1 Flow C...

Page 4: ...4 3 Change Decimal Point 8 TROUBLESHOOTING 8 1 Common Conditions 8 2 Troubleshooting Guide 8 3 Technical Assistance 9 CALIBRATION CONVERSIONS FROM REFERENCE GASES APPENDIX 1 COMPONENT DIAGRAM APPENDIX...

Page 5: ...ase contact an OMEGA7 customer service representative and request a Return Authorization Number AR It is mandatory that any equipment returned for servicing be purged and neutral ized of any dangerous...

Page 6: ...t or other contaminant s Be sure to observe the direction of gas flow as indicated by the arrow on the front of the meter when connecting the gas system to be monitored Insert tubing into the compress...

Page 7: ...PLY VOLTAGE LIMIT MAY RESULT IN DEVICE PERMANENT DAMAGE IMPORTANT NOTES In general D Connector numbering patterns are standardized There are how ever some connectors with nonconforming patterns and th...

Page 8: ...as the LCD readout mounting Special lengths of remote extension wiring up to 9 5 feet 3 meters are available on request 2 3 2 Panel Mounting Readouts Another option for the FMA1700A 1800A Mass Flow Me...

Page 9: ...to the total flow through the transducer In order to sense the flow in the sensor tube heat flux is introduced at two sec tions of the sensor tube by means of precision wound heater sensor coils Heat...

Page 10: ...25 to 100 of full scale flow rate GAS PRESSURE 1000 psig 69 bars FMA1700A 1800A Series Max Flow 10 50 and 100 500 psig 34 5 bars FMA1700A 1800A Series Max Flow 200 500 and 1000 L min Optimum pressure...

Page 11: ...Series Max Flow 1000 L min 3 4 FNPT fittings or 3 4 compression fittings LCD DISPLAY 3 digit LCD maximum viewable digits 1999 0 5 inch high characters On FMA1700A 1800A aluminum or stainless steel mod...

Page 12: ...43 200 44 500 45 1000 TABLE I FMA1700A 1800A SERIES MAX FLOW 10 L min LOW FLOW MASS FLOW METER CODE mL min N2 02 0 to 10 04 0 to 20 06 0 to 50 08 0 to 100 10 0 to 200 12 0 to 500 CODE L min N2 14 0 t...

Page 13: ...seepage or leak occurs in to the meter 5 2 Flow Signal Output Readings Initiate a controlled gas flow after warm up The flow signal output can be viewed either on the LCD display remote panel meter di...

Page 14: ...enance required It is good practice however to keep the meter away from vibration hot or corrosive environments and excessive RF or magnetic interference If periodic calibrations are required they sho...

Page 15: ...adaptor using a 5 32 Allan wrench 5 Clean or replace each of the removed parts as necessary If alcohol is used for cleaning allow time for drying Inspect the flow path inside the transducer for any vi...

Page 16: ...support for Mass Flow Meters using precision calibrators under strictly controlled conditions NIST traceable calibrations are available Calibrations can also be per formed at customers site using avai...

Page 17: ...least four times better collective accuracy than that of the Mass Flow Meter to be calibrated be used Equipment required for calibration includes a flow calibration standard and a certified high sens...

Page 18: ...ulated screwdriver adjust the ZERO potentiometer R34 through the access window for 0 VDC or 4 mA respectively at zero flow 14 CAUTION The minimum voltage on 0 5 Vdc output can be in the range of 7 to...

Page 19: ...arily disconnect the gas source Using the multimeter and the insulated screwdriver adjust the ZERO potentiometer R34 through the access window for 0 VDC or 4 mA respectively at zero flow CAUTION The m...

Page 20: ...e signal Adjust the setting for potentiometer R38 by using the insu lated screwdriver through the access window until reading is within specification 7 3 7 75 Flow Adjustment Using the flow regulator...

Page 21: ...ired flow and pressure conditions for a given gas or a mix ture of gases It was carefully packed to prevent damage during shipment Should you feel that the instrument is not functioning properly pleas...

Page 22: ...700A 1800A transducer from power supply remove the shorting condition or check polarities fuse resets automatically flush clean or disassemble to remove impediments or replace flush clean or or return...

Page 23: ...mum is recommended Carefully remove the LCD from FMA1700A 1800A Remove the aluminum housing on the side of the connection cable Side the LCD assembly out of the aluminum housing Check connection for a...

Page 24: ...ns 2 and 3 of the D connector Make sure the load resistance is more than 1000 Ohm Return FMA1700A 1800A to factory for repair Check external connections to pins 8 and 9 of the D connector Make sure th...

Page 25: ...12 NOTE Calibration accuracy can be affected Return FMA1700A 1800A to factory for repair Flush clean or disassemble to remove impediments or replace the filter screen see 6 2 on page 11 NOTE Calibrat...

Page 26: ...rom FMA1700A 1800A Remove the aluminum housing on the side of the connection cable Side the LCD assembly out of the aluminum housing Reinstall jumper to the appropriate location on the 8 pin header bl...

Page 27: ...S FROM REFERENCE GASES The calibration conversion incorporates the K factor The K factor is derived from gas density and coefficient of specific heat For diatomic gases K Kgas Note in the above relati...

Page 28: ...24 APPENDIX 1 COMPONENTS DIAGRAM FMA1700A 1800A METERING PC BOARD...

Page 29: ...uoride BrF3 3855 1161 6 108 Bromotrifluoromethane Freon 13 B1 CBrF3 3697 1113 6 644 1 3 Butadiene C4H6 3224 3514 2 413 Butane C4H10 2631 4007 2 593 1 Butene C4H8 2994 3648 2 503 2 Butene C4H8 CIS 324...

Page 30: ...55 Ethyl Acetylene C4H6 3225 3513 2 413 Ethyl Chloride C2H5Cl 3891 244 2 879 Ethylene C2H4 60 365 1 251 Ethylene Oxide C2H4O 5191 268 1 965 Fluorine F2 9784 1873 1 695 Fluoroform Freon 23 CHF3 4967 17...

Page 31: ...mide CH3Br 5835 1106 4 236 Methyl Chloride CH3Cl 6299 1926 2 253 Methyl Fluoride CH3F 68 3221 1 518 Methyl Mercaptan CH3SH 5180 2459 2 146 Methyl Trichlorosilane CH3 SiCl3 2499 164 6 669 Molybdenum He...

Page 32: ...35 1592 6 516 Sulfuryl Fluoride SO2F2 3883 1543 4 562 Tetrafluoroethane Forane 134A CF3CH2F 5096 127 4 224 Tetrafluorohydrazine N2F4 3237 182 4 64 Trichlorofluoromethane Freon 11 CCl3F 3287 1357 6 129...

Page 33: ...A B C D E F G H J 10 L min Tube O Diameter 5 72 145 3 1 00 25 4 1 13 28 6 3 09 78 6 5 10 129 6 0 5 12 7 4 61 117 1 0 16 4 0 0 16 4 0 50 L min Tube O Diameter 6 10 154 9 1 38 34 9 1 25 31 8 4 13 104 8...

Page 34: ...169 8 8 81 223 7 0 88 22 2 5 62 142 6 4 69 119 0 1 39 35 3 1 00 25 4 0 18 4 6 10 24 x 0 25 500 L min Tube O Diameter 7 64 194 0 3 00 76 2 3 00 76 2 7 25 184 2 9 65 245 1 1 5 38 1 6 53 165 8 6 75 171 5...

Page 35: ...31 PARTS OF THE FLOW METER LEFT AND RIGHT VIEWS FRONT VIEW...

Page 36: ...32...

Page 37: ...33...

Page 38: ...34...

Page 39: ...therwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold...

Page 40: ...Air Velocity Indicators Turbine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps I...

Reviews: