BASIC FUNCTIONS
7.0 Basic Functions
7.1.1 In standby mode, the current date, time and
ambient temperature will be displayed. The
date display range lies between 2001-2099
(Gregorian calendar), the time display can be
set to be displayed in either 12 or 24-hour
format. The starting time when instrument is reset is
set to be 01.01.2008. An announcement of the time and
the ambient temperature can be made by pressing the
[TALK/OK]
button. The blue background display light can
be engaged using the
[]
button.
7.1.2 The range of measurements, and of display, for body
temperature, lies between 32.0 °C and 42.9°C.
7.1.3 A maximum of 20 body temperature readings will be
saved. Each stored reading has a maximal save time of
20 days. The body temperature will be saved along with
the indication of type of measurement made (forehead
or ear) and details as to the time and date. The saved data
can be read out by pressing the
[TALK/OK]
button.
7.1.4 Either a forehead or an aural (ear) measurement can be
carried out with the forehead-and-ear thermometer.
When the body temperature measurement is successfully
completed, two short signal tones will be heard and,
depending on set-up at the time, the reading will be
read out by the audio output function. When an body
temperature measurement is being made, the light can
be switched on by pressing the
[]
button.
2634-09DST_M_GB_GebrAnw_Stirn_Ohrthermo_2.indd 12