12
Input Type Selection
46
12.3 INPUT TYPE: (DC VOLTAGE/PROCESS VALUE)
ENTER INPUT TYPE MENU:
Press ‘
MENU
’
1) Display shows
"
INPUt
"
Input Menu.
Press ‘
RESET/ENTER
’
2) Display flashes
"
VOLt
"
(default) or previous stored input type.
Press ‘
䊱
/MAX
’
3) Scroll through the available following input type selection list. Display
flashes:
"
VOLt
"
(DC Voltage)
"
CURRNt
"
(DC Current)
"
bRIdGE
"
(Bridge)
"
POt
"
(Potentiometer)
"
tC
"
(Thermocouple)
"
Rtd
"
(RTD)
If the displayed input type is
"
VOLt
"
, press ‘
RESET/ENTER
’ to skip to Step 6
(
"
VOLt
"
stops flashing).
VOLt
“VOLt”
DC VOLTAGE SUBMENU:
Press ‘
䊱
/MAX
’
4) Scroll through the available selection to
"
VOLt
"
(flashing).
Press ‘
RESET/ENTER
’
5) Display shows
"
VOLt
"
again, but not flashing.
Press ‘
RESET/ENTER
’
6) Display flashes
"
UN IPOL
"
unipolar (default) or previously stored
selection of voltage input type
"
bI POL
"
bipolar.
You must choose bi-Polar or uni-Polar before selecting Bridge or
Potentiometer
Press ‘
䊱
/MAX
’
7) Scroll through the 2 available voltage input types to the selection of
your choice.
Press ‘
RESET/ENTER
’
8) Display flashes
"
100 MV
"
(default) or previous stored input voltage
range selection. If the display matches your choice of Voltage range. Go
to Step 10.
Press ‘
䊱
/MAX
’
9) Scroll through the available voltage ranges to the selection of your
choice (see below Table for Voltage Input types/Ranges selection).
Press ‘
RESET/ENTER
’
10) Display shows
"
StOREd
"
stored message momentarily and then
advances to the
"
RdG.CNF
"
Reading Configuration Menu.
Display
Voltage types Voltage Ranges
UN IPOL
100 MV
unipolar
100 milliVolt DC
1 V
unipolar
1 Volt DC
10 V
unipolar
10 Volt DC
100 V
unipolar
100 Volt DC
bI POL
50 MV
bipolar
± 50 milliVolt DC
500 MV
bipolar
± 500 milliVolt DC
5 V
bipolar
± 5 Volt DC
50 V
bipolar
± 50 Volt DC