Configuring The Meter
26
4.3.1 Scaling with Known Loads (On-Line Calibration) (Continued)
11.
Apply a known load equal to approximately 100% of the transducer range.
12.
Press the
∂
TARE
button again. The meter shows the last stored value for Input 2.
13.
Press the
∂
TARE
button once more. The meter shows the actual signal being
received.
14.
Press the
MENU
button to store Input 2 value. The meter shows
RD!2
(Read 2).
RD!2
(Read 2) is the desired display reading at input 2.
15.
Press the
∂
TARE
button. The meter shows the last stored value for Read 2.
16.
Press the
ß
MAX
button to change the value of your digits.
17.
Press the
∂
TARE
button to scroll horizontally to the next digit.
18.
Press the
MENU
button to store value as
RD!2
(Read 2). The meter momentarily
shows
STRD
, followed by
RD.CF
. Meter scaling is now complete.