Configuring The Meter
4
4.4.1 Enabling or Disabling the Front-Panel Tare
To enable or disable the front-panel tare, follow these steps:
1.
Press the
MENU
button until
RD.CF
displays.
2.
Press the
∂
TARE
button. The meter shows one of the following:
•
R.1=T
(Tare enabled)
(Default)
•
R.1=N
(Tare disabled)
3.
Press the
ß
MAX
button to view last stored selection. Press the
ß
MAX
button to
toggle between selections.
4.
Press the
∂
TARE
button to select input resolution or press the
MENU
button to
store your selections.
STRD
momentarily displays, followed by
COLR
menu.
4.4.2 Setting Input Resolution
To set the input resolution of your meter, follow these steps:
1.
Press the
MENU
button until
RD.CF
displays, then press the
∂
TARE
button twice.
or
Press the
∂
TARE
button from
R.1
.
One of the following displays (default is
R.2=4
):
R.2=4
= 10 µV for Unipolar inputs. 25 µV for Bipolar inputs
R.2=0
= 1 µV for Unipolar inputs. 5 µV for Bipolar inputs.
R.2=1
= 2 µV for Unipolar inputs. 10 µV for Bipolar inputs
R.2=2
= 3 µV for Unipolar inputs. 15 µV for Bipolar inputs.
R.2=3
= 5 µV for Unipolar inputs. 25 µV for Bipolar inputs
Example:
3 µV resolution means that if you input 0-30 mV, at 30 mV the display shows
9999
.
2.
Press the
ß
MAX
button to scroll through available selections.
3.
Press the
∂
TARE
button to display the filtered/unfiltered signal input or press the
MENU
button to store your selections.
STRD
momentarily displays, followed by
COLR
menu.