12 

MANUAL DATALOGGING  (MAX 199 POINTS) 

1. Set the sampling rate to “0” (Refer to “

Setting the datalogger sampling rate

”).

2. In the manual mode, data is logged when press and hold 

ADJ

 button and screen shows recorded points “00X” in the

temp. block with icon “

MEM

” flash in few seconds.

e.g.

 Recorded 1st point, then bottom screen shows “001”.

3. Long press 

CAL

 button to clear data (Removed MANUAL.csv), screen shows “

CLr

”.

NOTE 1: 

While screen shows “

Err

” by long pressing 

CAL 

button, it stands for no data can be cleared or

   SD card isn’t being inserted. 

NOTE 2:

 Once clear the data by long press 

CAL

, there is no way to recover the data. If you want to   

   keep the previous data, rename the file “

MANUAL.csv

” in 

/

DOH-10

/ MANUAL.csv

 is required. 

4.

 

Data directory in SD card : 

/

DOH-10

/ MANUAL.csv

NOTE :

 When manual data records full (199 points), logging will continue, but with new data overwriting

  old. If you want to keep the previous data, rename the file “

MANUAL.csv

” in 

/

DOH-10

/  

  

MANUAL.csv

 is required. 

TRANSFERRING SD DATA TO PC 

a)

Remove SD card from the meter.

b)

Insert SD card directly into a PC SD card slot or use a SD card reader.

c)

Open the saved documents (CSV.) (Data stored) in the folder from PC.

d)

File name /Product number/ Sample rate/ Recording point/ Start recording time/ End recording time/ Recording
date/time /Recording parameters will be shown in the CSV. file.

e)

Data show “-49” stands for no measured value during recording period.

Manual Datalogging

 A

uto

 Datalogging

 

Summary of Contents for DOH-10 Series

Page 1: ...ser sGuide Shop online at omega com e mail info omega com For latest product manuals omega com en us pdf manuals DOH 10 Series Handheld Dissolved Oxygen Meter Kit with Optional SD Card Data Logger TM...

Page 2: ...78 USA Canada only Engineering Service 1 800 872 9436 USA Canada only Tel 203 359 1660 Fax 203 359 7700 e mail info omega com For Other Locations Visit omega com worldwide The information contained in...

Page 3: ...l testing lakes streams oceans Water Wastewater treatment Wine production Tripod receptacle mountable design for long time monitoring purposes 1 Meter 2 Battery AAA x 3 pcs 3 Electrode x 1pcs DO Galva...

Page 4: ...g mode UNIT Short press to change temperature unit Long press to enter temperature electrode type selection Short press to select NTC Negative Temperature Coefficient NOT no remote temperature electro...

Page 5: ...rwise LCD shows erratic readings and leakage may occur Warranty is void if not follow the notices Note Remove batteries when not use Power off switch When the meter is turned off the internal CPU does...

Page 6: ...used B DO electrode installation NOTE DO NOT touch the sensitive membrane when using the electrode or replacing the membrane cap for the sweat and grease will affect the quality of membrane and decre...

Page 7: ...Push the membrane cap downward gently until it won t lose 7 Fill the membrane cap with the DO electrolyte solution 0 5M NaOH provided Some solution will be vented from the overflow port This is norma...

Page 8: ...n lab wipe C DO calibration C 1 or C 2 or C 3 There are three ways to do the DO calibration for common and convenient way follow c 1 For more accurate measurement follow c 2 and c 3 with necessary dev...

Page 9: ...r and wait for the reading to stabilize press and hold MODE CAL buttons to enter calibration mode screen shows DO 100 then press and hold SET button to save and screen shows SA to complete calibration...

Page 10: ...ce is constantly replenished A moving stream will provide adequate circulation Freeze the current readings of DO and temperature readings by press HOLD button then icon Hold will appear on the top lef...

Page 11: ...2 5 08 15 0 10 08 9 54 9 03 8 54 8 08 7 64 41 0 6 31 6 03 5 75 5 49 5 24 5 01 16 0 9 87 9 34 8 84 8 37 7 92 7 5 42 0 6 21 5 93 5 67 5 41 5 17 4 93 17 0 9 67 9 15 8 67 8 21 7 77 7 36 43 0 6 12 5 84 5 5...

Page 12: ...ust connected to your computer Right click on your SD card to bring up the right click menu options Select Format Keep Capacity and Allocation unit size set to be default c Select the file system This...

Page 13: ...serted 4 When an SD card is used for the first time a folder is created on the card and named with the model number Under the MODEL number folder the MODEL number and AUTO YEAR folder will be automati...

Page 14: ...ou want to keep the previous data rename the file MANUAL csv in DOH 10 MANUAL csv is required 4 Data directory in SD card DOH 10 MANUAL csv NOTE When manual data records full 199 points logging will c...

Page 15: ...rodes When not in use Long term For long term storage or to take the electrode out of service disconnect the electrode from the meter Disassemble the electrode membrane cap Rinse the anode cathode and...

Page 16: ...se shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold by OME...

Reviews: