Omega DOH-10 Series User Manual Download Page 9

SALINITY CORRECTION 

BAROMETRIC PRESSURE SETTING: 

C-2) Zero Dissolved Oxygen Calibration: 

(Laboratory calibration with Na

2

SO

powder)

NOTE: 

Generally need to do zero oxygen calibration in the situation of replacing 

new electrode, replacing membrane cap, and long time without using. To do zero dissolved 
oxygen calibration by the following steps: 
1) Connect the electrode to the meter.
2) Use the beaker by dissolving approximately 10 grams of Na

2

SO

in 500 ml of

distilled water.

3) Place the electrode in the Na

2

SO

solution and wait for the reading to stabilize,

press and hold 

MODE+CAL 

buttons to enter calibration mode, press 

MODE+ADJ+UNIT 

buttons

again to enter “

DO %0.0

” mode, then press and hold 

SET 

button to save and screen shows “

SA” 

to complete

calibration.
Press and hold 

MODE+CAL 

buttons, screen shows 

“ESC” 

momentarily and return to normal measurement

mode

C-3) 100% Air-Saturated Water Calibration: 

(Laboratory calibration with stirrer)

1) Connect the electrode to the meter.
2) Pour 100ml deionized water into a 150mL beaker. Use an air bubbler or some

type of aerator to bubble air through the water while stirring for 

20 

minutes until

the water becomes completely saturated with air.

3) Place the electrode in the air-saturated water and wait for the reading to stabilize,

press and hold 

MODE+CAL 

buttons to enter calibration mode, screen shows “

DO %100

”, then press and hold 

SET

button to save and screen shows “

SA” 

to complete calibration. Press and hold 

MODE+CAL 

buttons, screen shows

“ESC” 

momentarily and return to normal measurement mode.

Due to the solution contains significant concentration of salinity affects the read out values of DO. 
Thus, to correct salinity value is required in order to get the accurate DO reading. 

(Refer to 

page 8

CHART 1

. for reference). Use a salinity meter to get the salt concentration reading. 

1) To enter the known salinity value by long press 

SET 

button and the screen shows “

SAL

”.

HOLD: 

↑to  increase 

ADJ: 

↓to  decrease

SET: 

← to left digit 

CAL: 

→ to right 

digit 

Adjustable range is 0 to 45.2 ppt. 

2) After setting completed, press and hold 

MODE 

button to save setting, screen shows “

SA

to complete setting. To escape the setting by press and hold 

SET 

button, screen shows

ESC

” and return to normal measurement mode.

If you are performing measurements at an altitude different from sea level 760 mmhg 
(default value). It is important to enter correct barometric pressure as barometric 
pressure affects DO values.  (Refer to 

page 9

CHART 2

. for reference) 

1) To enter the known pressure value by long press 

SET 

button and the screen shows “

SAL

”.

Then, press 

MODE 

button again to switch to the pressure setting mode, screen

shows “

P

”. 

HOLD: 

↑to

 increase 

ADJ: 

↓to

 decrease 

Summary of Contents for DOH-10 Series

Page 1: ...ser sGuide Shop online at omega com e mail info omega com For latest product manuals omega com en us pdf manuals DOH 10 Series Handheld Dissolved Oxygen Meter Kit with Optional SD Card Data Logger TM...

Page 2: ...78 USA Canada only Engineering Service 1 800 872 9436 USA Canada only Tel 203 359 1660 Fax 203 359 7700 e mail info omega com For Other Locations Visit omega com worldwide The information contained in...

Page 3: ...l testing lakes streams oceans Water Wastewater treatment Wine production Tripod receptacle mountable design for long time monitoring purposes 1 Meter 2 Battery AAA x 3 pcs 3 Electrode x 1pcs DO Galva...

Page 4: ...g mode UNIT Short press to change temperature unit Long press to enter temperature electrode type selection Short press to select NTC Negative Temperature Coefficient NOT no remote temperature electro...

Page 5: ...rwise LCD shows erratic readings and leakage may occur Warranty is void if not follow the notices Note Remove batteries when not use Power off switch When the meter is turned off the internal CPU does...

Page 6: ...used B DO electrode installation NOTE DO NOT touch the sensitive membrane when using the electrode or replacing the membrane cap for the sweat and grease will affect the quality of membrane and decre...

Page 7: ...Push the membrane cap downward gently until it won t lose 7 Fill the membrane cap with the DO electrolyte solution 0 5M NaOH provided Some solution will be vented from the overflow port This is norma...

Page 8: ...n lab wipe C DO calibration C 1 or C 2 or C 3 There are three ways to do the DO calibration for common and convenient way follow c 1 For more accurate measurement follow c 2 and c 3 with necessary dev...

Page 9: ...r and wait for the reading to stabilize press and hold MODE CAL buttons to enter calibration mode screen shows DO 100 then press and hold SET button to save and screen shows SA to complete calibration...

Page 10: ...ce is constantly replenished A moving stream will provide adequate circulation Freeze the current readings of DO and temperature readings by press HOLD button then icon Hold will appear on the top lef...

Page 11: ...2 5 08 15 0 10 08 9 54 9 03 8 54 8 08 7 64 41 0 6 31 6 03 5 75 5 49 5 24 5 01 16 0 9 87 9 34 8 84 8 37 7 92 7 5 42 0 6 21 5 93 5 67 5 41 5 17 4 93 17 0 9 67 9 15 8 67 8 21 7 77 7 36 43 0 6 12 5 84 5 5...

Page 12: ...ust connected to your computer Right click on your SD card to bring up the right click menu options Select Format Keep Capacity and Allocation unit size set to be default c Select the file system This...

Page 13: ...serted 4 When an SD card is used for the first time a folder is created on the card and named with the model number Under the MODEL number folder the MODEL number and AUTO YEAR folder will be automati...

Page 14: ...ou want to keep the previous data rename the file MANUAL csv in DOH 10 MANUAL csv is required 4 Data directory in SD card DOH 10 MANUAL csv NOTE When manual data records full 199 points logging will c...

Page 15: ...rodes When not in use Long term For long term storage or to take the electrode out of service disconnect the electrode from the meter Disassemble the electrode membrane cap Rinse the anode cathode and...

Page 16: ...se shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold by OME...

Reviews: