BUTTON DESCRIPTION:
ELECTRODE INSTALLATION (BNC connector):
Temperature
0~90
℃
Temp. accuracy
±0.5
℃
DO (Dissolved Oxygen) Electrode
Measurement range
0~199.9% (In saturation); 0.0~20.0 mg/L
Accuracy
±2% of full scale + 1 digit
Resolution
0.1%, 0.1 mg/L
Calibration
100% Air-Saturated
Flow condition
0.3 mL/s
Dimension
12x120mm
Electrode body
ABS
Sensor type
Galvanic
ATC Temp. sensor probe
port
3.5 Ø mm diameter phone jack (10K ohm resistance)
Cable length
3 M
PWR
Power on (Press in one second) or power off (Press more than 2 seconds when
operation)
SET
Long press to enter/escape salinity/pressure setting.
Move to left digit. (Under setting mode).
Long press to save setting or calibrate the reading.
CAL
Move to right digit. (Under setting mode).
MODE
Short press to change DO unit (mg/L or %).
Short press to enter pressure setting. (Under salinity setting mode.)
UNIT
Short press to change temperature unit
℃
/
℉
.
Long press to enter temperature electrode type selection.
Short press to select NTC: Negative Temperature Coefficient)/ NOT: no remote
temperature electrode.
Long press to save setting.
HOLD
Freeze current readings (
Hold
icon shows at top of LCD).
Increase value. (Under setting mode).
ADJ
Decrease value. (Under setting mode).
MODE+CAL
Long press to enter DO 100% or zero calibration mode.
(Na
2
SO
3
is required).
SET+UNIT
Recover to factory setting (under DO calibration mode).
HOLD+PWR
Disable Auto Power Off.
1.
Insert the DO electrode into the top of right hole. And insert the 3.5mm Ø diameter phone jack ATC sensor plug
into the middle hole.
2.
Hold the BNC connector in one hand; with the other, insert the braid into the center of the connector. Continuing
pushing the braid into the connector until it will not go in any farther.
Do this gently and slowly; do not bend the braid.
3.
Turn the male BNC connector in a clockwise direction, until you cannot turn it any more.
DO ELECTRODE SPECIFICATION: