Omega DOH-10 Series User Manual Download Page 13

11 

4. Re-power 

on

 the meter to back to normal measurement mode.

NOTE:

 To escape setting by turning 

off

 the meter without any change.

b) Setting the datalogger sampling rate

1. While meter is power 

on

, press and hold 

MODE

 button to enter setting.

2. Press 

HOLD

 button to increase the value ; press 

ADJ

 button to decrease the

value. 

3. Long press 

MODE

 button to save setting.

4. The available settings are: 2 sec, 5 sec, 10 sec, 15 sec, 30 sec, 60 sec, 120 sec, 300 sec, 600 sec, 900 sec, 1800 sec, and

1hr.

c)

Start datalogging

Warning:

 SD recording the selected temperature unit (

or

). If changing the temperature unit

during the datalogging sessions, the recorded data will be switched into the selected temperature unit.

1. After inserting the SD card, display will show icon “Logging” on the bottom of the screen.
2. Long press 

ADJ

 button to starts recording until icon “

Logging

” flashing on the bottom of screen.

3. When “

-Sd-

” disappear, SD stop to record data or SD card is not being inserted.

4. When an SD card is used for the first time ,a folder is created on the card and named with the

 model number

. Under

the MODEL number folder, the MODEL number and 

AUTO+YEAR

 folder will be automatically created. 

e.g.:

/

DOH-10

/AUTO2017/…..

 

5.

When starts datalogging, a new folder named 

M(month)/D(date)/H(hour)/M(minute)

 is created on the SD card in

 

the 

AUTO+YEAR

 folder. At the same time, a new spreadsheet document (CSV.) named M/D/H/M is also created

 

under its 

folder.

e.

g.: /

DOH-10

/AUTO2017/04051858/04051858.csv

6.

Each CSV. file can be stored up to 30,000 points.
Once 30,000 points are stored, a new file name will be auto created as 

M/D/H/M

 right after the last recording time.

Unless you interrupt recording, this process continues in the initial created M/D/H/M folder. 

e.g.: /

DOH-10

/AUTO2017/12261858/12262005.csv

NOTE1: 

Datalogging stopped when replacing the electrode or removing the SD card or resetting the

sampling rate.

NOTE2: 

When the recording is been stopped, a new folder will be created as M/D/H/M from the next datalogging.

NOTE3: 

When the recording year and model number is changed, the new folder will be also created

  accordingly. 

Month  Date 

Hour 

Minute 

Summary of Contents for DOH-10 Series

Page 1: ...ser sGuide Shop online at omega com e mail info omega com For latest product manuals omega com en us pdf manuals DOH 10 Series Handheld Dissolved Oxygen Meter Kit with Optional SD Card Data Logger TM...

Page 2: ...78 USA Canada only Engineering Service 1 800 872 9436 USA Canada only Tel 203 359 1660 Fax 203 359 7700 e mail info omega com For Other Locations Visit omega com worldwide The information contained in...

Page 3: ...l testing lakes streams oceans Water Wastewater treatment Wine production Tripod receptacle mountable design for long time monitoring purposes 1 Meter 2 Battery AAA x 3 pcs 3 Electrode x 1pcs DO Galva...

Page 4: ...g mode UNIT Short press to change temperature unit Long press to enter temperature electrode type selection Short press to select NTC Negative Temperature Coefficient NOT no remote temperature electro...

Page 5: ...rwise LCD shows erratic readings and leakage may occur Warranty is void if not follow the notices Note Remove batteries when not use Power off switch When the meter is turned off the internal CPU does...

Page 6: ...used B DO electrode installation NOTE DO NOT touch the sensitive membrane when using the electrode or replacing the membrane cap for the sweat and grease will affect the quality of membrane and decre...

Page 7: ...Push the membrane cap downward gently until it won t lose 7 Fill the membrane cap with the DO electrolyte solution 0 5M NaOH provided Some solution will be vented from the overflow port This is norma...

Page 8: ...n lab wipe C DO calibration C 1 or C 2 or C 3 There are three ways to do the DO calibration for common and convenient way follow c 1 For more accurate measurement follow c 2 and c 3 with necessary dev...

Page 9: ...r and wait for the reading to stabilize press and hold MODE CAL buttons to enter calibration mode screen shows DO 100 then press and hold SET button to save and screen shows SA to complete calibration...

Page 10: ...ce is constantly replenished A moving stream will provide adequate circulation Freeze the current readings of DO and temperature readings by press HOLD button then icon Hold will appear on the top lef...

Page 11: ...2 5 08 15 0 10 08 9 54 9 03 8 54 8 08 7 64 41 0 6 31 6 03 5 75 5 49 5 24 5 01 16 0 9 87 9 34 8 84 8 37 7 92 7 5 42 0 6 21 5 93 5 67 5 41 5 17 4 93 17 0 9 67 9 15 8 67 8 21 7 77 7 36 43 0 6 12 5 84 5 5...

Page 12: ...ust connected to your computer Right click on your SD card to bring up the right click menu options Select Format Keep Capacity and Allocation unit size set to be default c Select the file system This...

Page 13: ...serted 4 When an SD card is used for the first time a folder is created on the card and named with the model number Under the MODEL number folder the MODEL number and AUTO YEAR folder will be automati...

Page 14: ...ou want to keep the previous data rename the file MANUAL csv in DOH 10 MANUAL csv is required 4 Data directory in SD card DOH 10 MANUAL csv NOTE When manual data records full 199 points logging will c...

Page 15: ...rodes When not in use Long term For long term storage or to take the electrode out of service disconnect the electrode from the meter Disassemble the electrode membrane cap Rinse the anode cathode and...

Page 16: ...se shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold by OME...

Reviews: