SOURCE

Choose this function to provide a simulated thermocouple signal into controllers, temperature 
transmitters, indicators or any input devices that measure thermocouple sensors..

1) Disconnect the thermocouple sensor from the 

device to be calibrated.

2)  Select “

SOURCE

” with slide switch 

w

.

3)  Connect a thermocouple wire (matching the type 

of wire to sensor being simulated) with miniature 
male T/C connector to the inputs of the device 
being calibrated, making sure to check polarity. 
Millivolt outputs (without cold junction) may be 
connected with a copper (white) miniature T/C 
connector or the banana jacks with copper wire.

The output is adjusted in 0.1° (or 0.001 mV) 
increments by turning the knob 

e

 

while the HI/

SET/LO switch 

q

 is in the “HI”, “LO” or “SET” 

 

position. Press and turn the knob for faster dialing 
with 10° (or 0.100) increments.

Sourcing Thermocouple

Storing HI/SET/LO Outputs

STORING HI and LO Outputs

Choose this function to provide a simulated thermocouple signal into controllers, temperature 
transmitters, indicators or any other input device that measure thermocouple sensors..

1)  Store your high (SPAN) output temperature by moving the HI/SET/LO switch to the 

HI

 position 

and turn the 

e

 EZ-Dial knob until the desired temperature is on the display. Press and hold the  

knob until 

STORED

 appears to store the value. Release the knob.

2)  Store your low (ZERO) output temperature by moving the HI/SET/LO switch to the 

LO

 position 

and turn the 

e

 knob until the desired temperature is on the display. Press and hold the knob until 

STORED

 appears to store the value. Release the knob.

3) Instantly output your SPAN and ZERO temperature outputs by moving the HI/SET/LO switch 

between HI and LO. You may also select any third temperature output (such as mid-range) using 
the SET position on the HI/SET/LO switch.

HI

TYPE K °C

OFF

Push & Hold

to Store

Push & Turn

for

Fast Dialing

Double Click

for Options

THERMOCOUPLE SOURCE

SOURCE

HI

SET

LO

J‐T‐E‐K‐R‐S‐B‐N‐G‐C‐D‐L‐U‐P‐mV

TYPES

322.1

(Optional Boot Shown)

Summary of Contents for CL540ZA

Page 1: ...IN Shop online at CL540A CL540AZ CL540A and CL540AZ Thermocouple Source omega com e mail info omega com For latest product manuals omegamanual info User sGuide MADE IN Shop online at CL540A CL540AZ C...

Page 2: ......

Page 3: ...h the backlit display Calibrate directly in temperature C F Stop carrying around a millivolt source and thermocouple tables The PIECAL CL540A CL540ZA works with the thermocouples you use including typ...

Page 4: ...to the Omega CL540A CL540ZA Configuration Mode Use configuration to select C or F T C Type CL540ZA Only and Auto Off On Off CHANGING BATTERIES Low battery is indicated by BAT on the display Approximat...

Page 5: ...K R S B N G C D L J DIN U T DIN Platinel II and mV AUTO OFF If AUTO OFF is ON the unit will turn off after 30 minutes of inactivity to save battery life If AUTO OFF is OFF the unit will stay on until...

Page 6: ...nd LO Outputs Choose this function to provide a simulated thermocouple signal into controllers temperature transmitters indicators or any other input device that measure thermocouple sensors 1 Store y...

Page 7: ...charger four NiMh batteries AC DC cords Part 020 0103 Low Battery Low battery indication with nominal 1 hour of life left Protection against misconnection Over voltage protection to 60 V dc rated for...

Page 8: ...o 200 0 0 6 400 0 to 328 0 1 1 Chromel Constantan Jacket Purple Red Purple 200 0 to 100 0 0 4 328 0 to 148 0 0 7 100 0 to 850 0 0 3 148 0 to 1562 0 0 5 850 0 to 1000 0 0 4 1562 0 to 1832 0 0 7 R 13 3...

Page 9: ...0 1 6 2100 0 to 2320 0 1 1 3812 0 to 4208 0 2 0 D 1 0 to 50 0 0 8 30 0 to 122 0 1 4 W3 Re W25 Re Jacket White Red White Yellow 50 0 to 1400 0 0 6 122 0 to 2552 0 1 3 1400 0 to 1800 0 0 7 2552 0 to 32...

Page 10: ...ator Clips Optional Accessories Part Number Rubber Boot HH500A RB Small Carrying Case fits unit with or without boot SC HH500 Ni MH 1 Hour Charger with 4 Ni MH AA Batteries 020 0103 100 120 V AC input...

Page 11: ......

Page 12: ...ication used on humans or misused in any way OMEGA assumes no responsibility as set forth in our basic WARRANTY DISCLAIMER language and additionally purchaser will indemnify OMEGA and hold OMEGA harml...

Reviews: