background image

Page 3

Configuration

Configure the Calibrator

Move 

POWER SWITCH to 

“SOURCE”

 or 

“READ”

Setup

Double click the 

e

 DIAL KNOB at any time the 

unit is on and the following displays will appear 
for 15 seconds:

Turn the 

e

 DIAL KNOB to move through the 

menu. Press the 

e

 DIAL KNOB to toggle 

between OFF and ON or to scroll through the 
settings. 

MODEL 54#B  V#.##

DOUBLE CLICK

EZ-DIAL KNOB

FOR CONFIGURATION

>  EXIT   

 

 

      15

 

TEMP UNITS        °C

 

T/C   

 

 

       K

 

AUTO OFF   

     ON 

EXIT MENU

 - exits this menu immediately and 

saves any changes. Menu will automatically exit 
after 15 seconds of inactivity (countdown timer is 
displayed).

TEMP UNITS

 - pressing the knob will toggle 

between °C and °F.

T/C

 - 

CL540A: pressing the knob will toggle between 
the factory configured T/C Type and mV.

CL540ZA: pressing the knob will cycle through 
T/C types J, T, E, K, R, S, B, N, G, C, D, L 
(J-DIN), U (T-DIN), Platinel II and mV.

AUTO OFF

 - If AUTO OFF is ON, the unit will 

turn off after 30 minutes of inactivity to save 
battery life. If AUTO OFF is OFF the unit will stay 
on until the POWER SWITCH is moved to the off 
position.

Note:

 All settings are remembered even with the 

power off. Removing the batteries resets the 
values to factory defaults.

Simulating thermocouples requires the use of 
thermocouple or extension grade thermocouple 
wire. 

Plug thermocouple wires into the female 
miniature thermocouple connector mounted in 
the top end of the housing.

The CL540A/CL540ZA has two banana jacks 
mounted in the top end of the housing. These 
are not temperature compensated and are to be 
used only for millivolt signals.

Connections

Push & Hold

to

Store/Ramp

Push & Turn

for

Fast Dialing

Double Click

for Options

HI

TYPE K °C

Thermocouple Wire

Device

Requiring

Calibration

OFF

THERMOCOUPLE SOURCE

SOURCE

HI

SET

LO

J‐T‐E‐K‐R‐S‐B‐N‐G‐C‐D‐L‐U‐P‐mV

TYPES

322.1

(Optional Boot Shown)

Summary of Contents for CL540ZA

Page 1: ...IN Shop online at CL540A CL540AZ CL540A and CL540AZ Thermocouple Source omega com e mail info omega com For latest product manuals omegamanual info User sGuide MADE IN Shop online at CL540A CL540AZ C...

Page 2: ......

Page 3: ...h the backlit display Calibrate directly in temperature C F Stop carrying around a millivolt source and thermocouple tables The PIECAL CL540A CL540ZA works with the thermocouples you use including typ...

Page 4: ...to the Omega CL540A CL540ZA Configuration Mode Use configuration to select C or F T C Type CL540ZA Only and Auto Off On Off CHANGING BATTERIES Low battery is indicated by BAT on the display Approximat...

Page 5: ...K R S B N G C D L J DIN U T DIN Platinel II and mV AUTO OFF If AUTO OFF is ON the unit will turn off after 30 minutes of inactivity to save battery life If AUTO OFF is OFF the unit will stay on until...

Page 6: ...nd LO Outputs Choose this function to provide a simulated thermocouple signal into controllers temperature transmitters indicators or any other input device that measure thermocouple sensors 1 Store y...

Page 7: ...charger four NiMh batteries AC DC cords Part 020 0103 Low Battery Low battery indication with nominal 1 hour of life left Protection against misconnection Over voltage protection to 60 V dc rated for...

Page 8: ...o 200 0 0 6 400 0 to 328 0 1 1 Chromel Constantan Jacket Purple Red Purple 200 0 to 100 0 0 4 328 0 to 148 0 0 7 100 0 to 850 0 0 3 148 0 to 1562 0 0 5 850 0 to 1000 0 0 4 1562 0 to 1832 0 0 7 R 13 3...

Page 9: ...0 1 6 2100 0 to 2320 0 1 1 3812 0 to 4208 0 2 0 D 1 0 to 50 0 0 8 30 0 to 122 0 1 4 W3 Re W25 Re Jacket White Red White Yellow 50 0 to 1400 0 0 6 122 0 to 2552 0 1 3 1400 0 to 1800 0 0 7 2552 0 to 32...

Page 10: ...ator Clips Optional Accessories Part Number Rubber Boot HH500A RB Small Carrying Case fits unit with or without boot SC HH500 Ni MH 1 Hour Charger with 4 Ni MH AA Batteries 020 0103 100 120 V AC input...

Page 11: ......

Page 12: ...ication used on humans or misused in any way OMEGA assumes no responsibility as set forth in our basic WARRANTY DISCLAIMER language and additionally purchaser will indemnify OMEGA and hold OMEGA harml...