background image

Page 5

General

Accuracy

±(0.015% of R 0.009 mV)

Cold Junction Compensation

± 0.45°F (±0.25 °C)

Millivolt Range

-13.000 to 80.000 mV

Operating Temperature  Range

-25 to 60 °C (-10 to 140 °F)

Temperature Drift

≤ 50 ppm of range (includes mV and Cold Junction)

Relative Humidity Range

10 % ≤RH ≤90 % (0 to 35 °C), Non-condensing
10 % ≤RH≤ 70 % (35 to 60 °C), Non-condensing

Size
       With Boot

4.96 x 2.73 x 1.79 inches, 126 x 69 x 45 mm (L x W x H)
5.67 x 3.06 x 2.05 inches, 144 x 78 x 52 mm (L x W x H)

Weight
      With Boot

8.4 ounces, 0.24 kg (including batteries)
11 ounces, 0.32 kg (including batteries)

Batteries

Four “AA” Alkaline 1.5V (LR6)

Battery Life

50 Hours

Optional NiMh Rechargeable battery kit

120 VAC for North America Only; charger,  four NiMh 

batteries, AC & DC cords [Part # 020-0103]

Low Battery

Low battery indication with nominal 1 hour of life left

Protection against misconnection

Over-voltage protection to 60 V dc (rated for 30 seconds)

Display

High contrast graphic liquid crystal display. LED backlighting 

for use in low lit areas.

Source

Output Impedance

< 0.3 Ohms

Source Current

> 20 mA (drives 80 mV into 10 Ohms)

Noise

≤ 4 microvolts p-p for frequencies of 10 Hz or below

Specifications

(Unless otherwise indicated all specifications are rated from a nominal 23 °C,

70 % RH for 1 year from calibration)

Additional Information 

This product is calibrated on equipment traceable to NIST and includes a Certificate of Calibration. 

Test Data is available for an additional charge.

OMEGA ENGINEERING, INC. recommends a calibration interval of one year. Contact your local 

representative for recalibration and repair services.

Summary of Contents for CL540ZA

Page 1: ...IN Shop online at CL540A CL540AZ CL540A and CL540AZ Thermocouple Source omega com e mail info omega com For latest product manuals omegamanual info User sGuide MADE IN Shop online at CL540A CL540AZ C...

Page 2: ......

Page 3: ...h the backlit display Calibrate directly in temperature C F Stop carrying around a millivolt source and thermocouple tables The PIECAL CL540A CL540ZA works with the thermocouples you use including typ...

Page 4: ...to the Omega CL540A CL540ZA Configuration Mode Use configuration to select C or F T C Type CL540ZA Only and Auto Off On Off CHANGING BATTERIES Low battery is indicated by BAT on the display Approximat...

Page 5: ...K R S B N G C D L J DIN U T DIN Platinel II and mV AUTO OFF If AUTO OFF is ON the unit will turn off after 30 minutes of inactivity to save battery life If AUTO OFF is OFF the unit will stay on until...

Page 6: ...nd LO Outputs Choose this function to provide a simulated thermocouple signal into controllers temperature transmitters indicators or any other input device that measure thermocouple sensors 1 Store y...

Page 7: ...charger four NiMh batteries AC DC cords Part 020 0103 Low Battery Low battery indication with nominal 1 hour of life left Protection against misconnection Over voltage protection to 60 V dc rated for...

Page 8: ...o 200 0 0 6 400 0 to 328 0 1 1 Chromel Constantan Jacket Purple Red Purple 200 0 to 100 0 0 4 328 0 to 148 0 0 7 100 0 to 850 0 0 3 148 0 to 1562 0 0 5 850 0 to 1000 0 0 4 1562 0 to 1832 0 0 7 R 13 3...

Page 9: ...0 1 6 2100 0 to 2320 0 1 1 3812 0 to 4208 0 2 0 D 1 0 to 50 0 0 8 30 0 to 122 0 1 4 W3 Re W25 Re Jacket White Red White Yellow 50 0 to 1400 0 0 6 122 0 to 2552 0 1 3 1400 0 to 1800 0 0 7 2552 0 to 32...

Page 10: ...ator Clips Optional Accessories Part Number Rubber Boot HH500A RB Small Carrying Case fits unit with or without boot SC HH500 Ni MH 1 Hour Charger with 4 Ni MH AA Batteries 020 0103 100 120 V AC input...

Page 11: ......

Page 12: ...ication used on humans or misused in any way OMEGA assumes no responsibility as set forth in our basic WARRANTY DISCLAIMER language and additionally purchaser will indemnify OMEGA and hold OMEGA harml...

Reviews: