2.0 Operation Procedure 

2.1 Drawing of Unit 

 

 

 

Actual Dimensions: 195mm(H) x 92mm(W) x 53mm(D)

 

2.2 Descriptions of Buttons and Switches 

 

/

 

Buttons 

Temperature  is  displayed  in  either  degrees  Celsius(°C)  or  degrees 
Fahrenheit(°F).  When  the  thermometer  is  turned  on,  it  is  set  to  the 
temperature scale that was in use when the thermometer was last turned off. 
To change the temperature scale, press the °C or °F  key. 
 

  Button 

Press  “

”  button  to  toggle  the  on  and  off  backlight.  The  backlight  will 

switch-off automatically after 37 seconds. 
 

  Button (HOLD Mode) 

Press the HOLD key to enter the Data Hold mode, the “HOLD” annunciator is 
displayed.  When  HOLD  mode  is  selected,  the  thermometer  will  hold  the 
present reading and stop all further measurements. 
Pressing  the  HOLD  key  again  will  cancel  HOLD  mode  causing  the 
thermometer to resume taking measurements. 
 

  Button (K/J/T/E Input Thermocouple Type Selection) 

The TYPE key allows for selection of J, K, E or T thermocouple types as either 
the  input  or  simulated  output.  To  select  thermocouple  type  press  type  key 
once. Unit will briefly show 1888 and beep and then the type of thermocouple 
selected  will  be  indicated  on  the  left  hand  side  of  the  display.  Continue 
pressing  the  type  key  to  step  through  the  thermocouple  types  until  desired 
type is indicated.

Summary of Contents for CL3512A

Page 1: ...zech Repubic Tel 420 69 6311627 FAX 420 69 6311114 e mail czech omega com France 9 rue Denis Papin 78190 Trappes Tel 33 130 621 400 FAX 33 130 699 120 Toll Free in France 0800 4 06342 e mail france omega com Germany Austria Daimlerstrasse 26 D 75392 Deckenpfronn Germany Tel 49 07056 3017 FAX 49 07056 8540 Toll Free in Germany 0800 82 66342 e mail germany omega com United Kingdom ISO 9002 Certified...

Page 2: ...1 1 Safety Information 1 2 Unpacking 2 0 Operation Procedure 2 1 Unit Drawing 2 2 Description of Buttons and Switches 2 3 Read Mode Procedure 2 4 Source Mode Procedure 3 0 Operator Maintenance 4 0 Specifications 5 0 Calibration Procedure ...

Page 3: ... not use this instrument when working voltages at the measurement surface over 24V AC or DC WARNING To avoid damage or burns do not make temperature measurement in microwave ovens CAUTION Repeated sharp flexing can break the thermocouple leads To prolong lead life avoid sharp bends in the leads especially near the connector 2 1 2 Unpacking Remove the packing list and verify that all equipment has ...

Page 4: ...Press the HOLD key to enter the Data Hold mode the HOLD annunciator is displayed When HOLD mode is selected the thermometer will hold the present reading and stop all further measurements Pressing the HOLD key again will cancel HOLD mode causing the thermometer to resume taking measurements Button K J T E Input Thermocouple Type Selection The TYPE key allows for selection of J K E or T thermocoupl...

Page 5: ...de to allow a particular simulated temperature to be selected once the output has been adjusted close to the desired temperature using the selector switch 6 2 3 Read Mode Procedure 1 Plug Thermocouple Sensor into input T1 and or T2 2 Turn selector switch to Input 3 Select proper thermocouple type J K T or E using type button 4 Press T1 button to read thermocouple connected to input T1 Press T2 but...

Page 6: ...sired simulated value Note There are 9 temperature range switch positions arranged from left to right representing low to high temperatures The display will read OL if the output voltage for the meter is higher than the maximum voltage for that particular thermocouple type For type T this will be in the highest 4 positions type K in the highest 3 positions type J in the highest 2 positions and typ...

Page 7: ...hermocouple Range K TYPE 0 1 C 200 C to 1372 C 328 F to 1999 F J TYPE 0 1 C 210 C to 1200 C 346 F to 1999 F T TYPE 0 1 C 200 C to 400 C 328 F to 752 F E TYPE 0 1 C 220 C to 1000 C 364 F to 1832 F Calibration Range 210 C to 1372 C 364 F to 1999 F Auto range 0 1 C 1 C 0 1 F 1 F Accuracy Accuracy is specified for operating temperatures over the range of18 C to 28 C 64 F to 82 F for 1 year not incl ud...

Page 8: ...Voltage betweenT1 and T2 during measurement 1volt 12 5 0 CL3152A Calibration Procedure Note The following calibration procedure should be performed only by qualified technicians who have access to the items as following items Equipment Thermocouple calibrator had better 10 times greater than the measured meter 1 Short CAL jumper and turn on the meter to INPUT range 2 Press key display was showed 3...

Page 9: ...13 ...

Page 10: ...DISCLAIMER language and additionally purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product s in such a manner RETURN REQUESTS INQUIRIES Direct all warranty and repair requests inquiries to the OMEGA Customer Service Department BEFORE RETURNING ANY PRODUCT S TO OMEGA PURCHASER MUST OBTAIN AN AUTHORIZED RETURN AR NUMBER F...

Reviews: