background image

 

11 

GENERAL 
Display:
 3½ digit liquid crystal display (LCD) with a maximum reading of1999. 
Polarity: Automatic, positive implied, negative polarity indication. 
Overrange: -OL is displayed. 
Zero: Automatic. 
Low  battery  indication:  The  “

”  is  displayed  when  the  battery  voltage 

drops below the operating level. 

Measurement rate: 1 times/second. 
Accuracy: Stated accuracy at 23°C±5°C, <75% R.H. 
Dimensions: 195mm(H) x 92mm(W) x 53mm(D). 
Weight: approx. 9 oz. (250g) including battery. 
Input Connector: Accepts standard miniature thermocouple connectors (flat 

blades spaced 7.9mm, center to center SMP type). 

Battery Life: 100 hours typical with carbon zinc battery. 
Auto power off: The meter key switch inactive for more than 90 minutes. 
Temperature Coefficient: 0.1 times the applicable accuracy specification per 

°C from 0°C to 18°C and 28°C to 50°C (32°F to 64°F 

and 82°F to 122°F). 

Input  Protection:  24V  dc  or  24V  ac  rms  maximum  input  voltage  on  any 

combination of input pins. 

Maximum  Differential  Common  Mode  Voltage  (Maximum  Voltage 
betweenT1 and T2 during measurement):
 1volt.

 

12 

5.0 CL3152A Calibration Procedure 

Note:  The  following  calibration  procedure  should  be  performed  only  by 

qualified technicians who have access to the items as following items: 

 
Equipment:  Thermocouple  calibrator  had  better  10  times  greater  than  the 
measured meter. 
 
1. Short CAL jumper and turn on the meter to INPUT range. 
2. Press “

” key display was showed 35°C / T1/ SET / K. 

3.  Input  35°C  /  K-type  to  the  T1,  then  press  “

”  key  and  wait  for  the 

display was showed 1300°C / T1 / SET / K. 

4. Input 1300°C / K-type to the  T1, then  press “

” key and wait for the 

display was showed 0°C / T1 / SET / K. 

5. Input 0°C/K-type to the T1, then  press “

” key and wait for the display 

was  showed  “-OL”  then  press  “

”  key  and  wait  for  the  display  was 

35°C / T2 / SET / K. 

6.  Input  35°C  /  K-type  to  the  T2,  then  press  “

”  key  and  wait  for  the 

display was showed 1300°C / T2 / SET / K. 

7. Input 1300°C / K-type to the  T2, then  press “

” key and wait for the 

display was showed Out / K / °C / SET / T2. 

8.  Knob  being  with  clockwise  rotation  2  range  and  two  knobs  of  lower  left 

corner are set to the position of the middle (horizontal). 

9.  Press  “

”  key  and  wait  for  the  meter  auto  power  off  then  remove 

jumper. 

10. The calibration was completed.

Summary of Contents for CL3512A

Page 1: ...zech Repubic Tel 420 69 6311627 FAX 420 69 6311114 e mail czech omega com France 9 rue Denis Papin 78190 Trappes Tel 33 130 621 400 FAX 33 130 699 120 Toll Free in France 0800 4 06342 e mail france omega com Germany Austria Daimlerstrasse 26 D 75392 Deckenpfronn Germany Tel 49 07056 3017 FAX 49 07056 8540 Toll Free in Germany 0800 82 66342 e mail germany omega com United Kingdom ISO 9002 Certified...

Page 2: ...1 1 Safety Information 1 2 Unpacking 2 0 Operation Procedure 2 1 Unit Drawing 2 2 Description of Buttons and Switches 2 3 Read Mode Procedure 2 4 Source Mode Procedure 3 0 Operator Maintenance 4 0 Specifications 5 0 Calibration Procedure ...

Page 3: ... not use this instrument when working voltages at the measurement surface over 24V AC or DC WARNING To avoid damage or burns do not make temperature measurement in microwave ovens CAUTION Repeated sharp flexing can break the thermocouple leads To prolong lead life avoid sharp bends in the leads especially near the connector 2 1 2 Unpacking Remove the packing list and verify that all equipment has ...

Page 4: ...Press the HOLD key to enter the Data Hold mode the HOLD annunciator is displayed When HOLD mode is selected the thermometer will hold the present reading and stop all further measurements Pressing the HOLD key again will cancel HOLD mode causing the thermometer to resume taking measurements Button K J T E Input Thermocouple Type Selection The TYPE key allows for selection of J K E or T thermocoupl...

Page 5: ...de to allow a particular simulated temperature to be selected once the output has been adjusted close to the desired temperature using the selector switch 6 2 3 Read Mode Procedure 1 Plug Thermocouple Sensor into input T1 and or T2 2 Turn selector switch to Input 3 Select proper thermocouple type J K T or E using type button 4 Press T1 button to read thermocouple connected to input T1 Press T2 but...

Page 6: ...sired simulated value Note There are 9 temperature range switch positions arranged from left to right representing low to high temperatures The display will read OL if the output voltage for the meter is higher than the maximum voltage for that particular thermocouple type For type T this will be in the highest 4 positions type K in the highest 3 positions type J in the highest 2 positions and typ...

Page 7: ...hermocouple Range K TYPE 0 1 C 200 C to 1372 C 328 F to 1999 F J TYPE 0 1 C 210 C to 1200 C 346 F to 1999 F T TYPE 0 1 C 200 C to 400 C 328 F to 752 F E TYPE 0 1 C 220 C to 1000 C 364 F to 1832 F Calibration Range 210 C to 1372 C 364 F to 1999 F Auto range 0 1 C 1 C 0 1 F 1 F Accuracy Accuracy is specified for operating temperatures over the range of18 C to 28 C 64 F to 82 F for 1 year not incl ud...

Page 8: ...Voltage betweenT1 and T2 during measurement 1volt 12 5 0 CL3152A Calibration Procedure Note The following calibration procedure should be performed only by qualified technicians who have access to the items as following items Equipment Thermocouple calibrator had better 10 times greater than the measured meter 1 Short CAL jumper and turn on the meter to INPUT range 2 Press key display was showed 3...

Page 9: ...13 ...

Page 10: ...DISCLAIMER language and additionally purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product s in such a manner RETURN REQUESTS INQUIRIES Direct all warranty and repair requests inquiries to the OMEGA Customer Service Department BEFORE RETURNING ANY PRODUCT S TO OMEGA PURCHASER MUST OBTAIN AN AUTHORIZED RETURN AR NUMBER F...