Section 6 - Specifications
Target Plate Temperature Range:
A 11°C to 260°C*
(A 20°F to 500°F)*
Accuracy:
±1.5°C (±2.75°F)
Display Resolution:
0.1°
Stability:
±0.15°C (±0.3°F)**
Probe Well Uniformity:
±0.15°C (±0.3°F)
Heat-Up Time:
from 38 to 260°C (100 to 500°F) (15 Minutes)
Cool-Down Time:
from 260° to 38°C (500 to 100°F) (70
Minutes)
Ambient Environmental Conditions
Temperature:
0 to 50°C (32 to 122°F)*
Humidity:
0 to 80% RH, non-condensing
Internal Control Sensor:
Platinum RTD, 100
Ω
, .00385, Class A
Power Requirements
Model CL1000A, B, C, D:
3.15 Amps @ 115 VAC~ (±10%)
50/60 Hz, 275W
Model CL1000A, B, C, D-230 VAC:
1.6 Amps @ 230 VAC~ (±10%) 50/60 Hz,
275W
Dimensions:
(127 x 56 x 155 mm) 5"w x 2.2"h x 6.1"d
Weight:
3.8 lbs. (1.7 Kg.)
Approvals:
CE (-230 VAC models only)
Installation Category II
* Maximum setpoint must be reduced when operating at ambient temperatures in
excess of 24°C (75°F) See Section 2.3
** With stable, correct nominal line voltage. (See Section 2.4)
CL1000 Series Specifications
6
6-1