CL1000 Series Installation
2
2-2
Figure 2. Maximum Setpoint Temperature Versus Ambient Temperature
An increase in ambient temperature of 0.6°C (1°F) above 24°C (75°F) reduces the maximum
allowable probe well temperature by 4°C (7°F).
2.3.2 Relative Humidity
Only operate your calibrator in a Relative Humidity environment of up to 80%
RH (Non-Condensing).
2.4 Power Connection and Requirements
Electrical connections and wiring should be performed only by suitably trained personnel.
Connect to a stable, earth grounded AC mains supply of the correct voltage for
the model calibrator you have. Use the power cord provided.
Calibrator power requirements
• 3.15 amps @ 115 VAC (±10%) 50/60 Hz.
• 1.6 amps @ 230 VAC (±10%) 50/60 Hz.
CAUTION:
NOTE:
Amb. Temp. Max. Setpoint
°C (°F)
°C (°F)
24.0 (75)
260 (500)
24.4 (76)
252 (486)
25.6 (78)
244 (472)
26.7 (80)
237 (458)
27.8 (82)
229 (444)
28.9 (84)
221 (430)
30.0 (86)
213 (416)
31.1 (88)
205 (402)
32.2 (90)
198 (388)
33.3 (92)
190 (374)