CONDUCTIVITY STANDARD SOLUTION (1413 µS/cm at 25°C)
※
Conductivity reading will be affected by the temperature.
℃
℉
µS/cm
℃
℉
µS/cm
℃
℉
µS/cm
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
32
33.8
35.6
37.4
39.2
41
42.8
44.6
46.4
48.2
50
51.8
53.6
55.4
57.2
59
60.8
776
799
822
846
870
894
918
943
968
992
1017
1043
1068
1094
1119
1145
1171
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
62.6
64.4
66.2
68
69.8
71.6
73.4
75.2
77
78.8
80.6
82.4
84.2
86
87.8
89.6
91.4
1198
1224
1251
1277
1304
1331
1358
1386
1413
1441
1468
1496
1524
1552
1580
1608
1636
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
93.2
95
96.8
98.6
100.4
102.2
104
105.8
107.6
109.4
111.2
113
114.8
116.6
118.4
120.2
122
1665
1693
1722
1751
1780
1808
1837
1866
1896
1925
1954
1983
2013
2042
2071
2101
2103
CONDUCTIVITY STANDARD SOLUTION (12880 µS/cm at 25°C)
※
Conductivity reading will be affected by the temperature.
℃
℉
µS/cm
℃
℉
µS/cm
5
10
15
16
17
18
19
20
21
41
50
59
60.8
62.6
64.4
66.2
68
69.8
8,216
9,326
10,439
10,684
10,929
11,174
11,419
11,664
11,909
22
23
24
25
30
35
40
45
50
71.6
73.4
75.2
77
86
95
104
113
122
12,153
12,398
12,643
12,880
14,112
15,392
16,678
18,024
19,338
ELECTRODE MAINTENANCE
Proper maintenance will ensure faster measurements, improve accuracy and extend the lifetime of
the electrode.
Cond./EC/ TDS /Salt electrode
Conductivity meters and cells should be calibrated to a standard conductivity solution. Selecting standards
is very important, you should always choose one that has the approximate conductivity
of the solution to be
measured.
A polarized or fouled electrode must be cleaned to renew the active surface of the cell.
7