In most situations, hot water with a mild liquid detergent is an effective cleanser. 
Acetone easily cleans most organic matter, and hypochlorous solutions will remove algae,bacteria, or molds. 
To prevent cell damage, abrasives or sharp objects should not be used to clean an electrode.   
A cotton swab also works well for cleaning. 

SD CARD DATA LOGGING 

After inserting the SD card, the “SD” icon will appear on the bottom of display. When the meter supplied by 

batteries, once the icon becomes          with SD icon disappeared, it indicates the power is not enough to 
activate SD function but other functions are still working, now we suggest the user to use adaptor to power the 
meter. Once the battery icon becomes 

, it means the power is very weak, and should be replaced the new 

batteries for operating the meters.      

SD Card Information

a)

Power the meter with AAA batteries or 9V AC/DC adaptor.

b)

Insert an SD card (8G supplied) into the SD card slot at the right side of the meter. The card must be
inserted with the front of the card (label side) facing toward front of the meter.
After inserting the SD card, icon “SD” will appear on the bottom of screen.

c)

If the SD card is being used for the first time it is recommended that the card be formatted.

SD Card Formatting

NOTE: Always confirm that the device is compatible with the SD, SDHC or SDXC memory card before 
formatting. 
WARNING:  Backup all your data before formatting. Formatting will erase all data on the memory device. 

a)

Open the Computer window.

Click the Start or Windows menu and select Computer (Windows Vista/7) or My Computer (Windows XP). For 
Windows 8 users, type “computer” and click the Computer icon in the Apps search results. For Windows 10, 
open the File Explorer. Then find “This PC”. 

b)

Find your SD card.

The removable drive that appears last in the "Devices with Removable Storage" list should be the SD card that 
you just connected to your computer. Right-click on your SD card to bring up the right-click menu options. Select 
Format. This will take you to the Format window. Keep “Capacity” and “Allocation unit size” set to default. 

c)

Select the file system.

 This is the way files are stored on the card. Different systems   
use different file structures. In order for the SD card to be read 
by any device, select FAT32 as the file system. This will enable it to 
be read by cameras, phones, printers, Windows, Mac, and Linux   
computers, and more. 

1.

Select Quick Format.

2.

Click “Start”.

3.

Once the formatting is complete, you can close the window.

8

Summary of Contents for CDH-SD11

Page 1: ...e mail info omega com For latest product manuals www omegamanual info Shop online at omega com User sGuide CDH SD11 Conductivity TDS and Salinity Meter with Real Time SD Card Data Logger TM 2 e Fo...

Page 2: ...pecifications without notice Servicing North America U S A Omega Engineering Inc Headquarters Toll Free 1 800 826 6342 USA Canada only Customer Service 1 800 622 2378 USA Canada only Engineering Servi...

Page 3: ...two years Electrode is one year SUPPLIED Meter Electrode x 1pc Cond EC TDS Salt Battery AAA x 3 pcs 1413uS buffer solution x1 9V adaptor Black carrying case 8G SD card SPECIFICATION Model CDH SD11 Par...

Page 4: ...stant K value Temp coefficient compensation tds K factor value mode Long press again to escape the K value setting Recovery to factory setting under calibration mode SET CAL Conductivity Calibration H...

Page 5: ...uto power off function by pressing HOLD and PWR button n momentarily shows on screen now meter is at non sleep mode then turns to normal measurement Meter default auto power off 6 FACTORY RESET a Cond...

Page 6: ...r 2 Power on the meter by pressing PWR button 3 First to enter the cell constant K value of the Cond electrode is required e g Refer to the label on the top handle of the probe Long press SET UNIT to...

Page 7: ...right digit Long press SET button to finish calibration screen will show SA The adjustable range is 1 to 7 4 Or to escape setting by long press SET UNIT the bottom of screen will show ESC in few seco...

Page 8: ...measurement except to change the display unit from uS mS to PPM PPT Display will show TDS values at the same time the left bottom display will show the temp value of the measured solution NOTE If the...

Page 9: ...808 1837 1866 1896 1925 1954 1983 2013 2042 2071 2101 2103 CONDUCTIVITY STANDARD SOLUTION 12880 S cm at 25 C Conductivity reading will be affected by the temperature S cm S cm 5 10 15 16 17 18 19 20 2...

Page 10: ...c If the SD card is being used for the first time it is recommended that the card be formatted SD Card Formatting NOTE Always confirm that the device is compatible with the SD SDHC or SDXC memory card...

Page 11: ...essions the recorded data will be switched into the selected temperature unit 1 Power on the meter by AAA batteries or 9V AC DC adaptor 2 After inserting the SD card display will show icon SD on the b...

Page 12: ...for no data can be cleared or SD card isn t being inserted NOTE Once clear the data by long press CAL there is no way to recover the data If you want to keep the previous data rename the file MANUAL c...

Page 13: ...ensor type or adjust temperature manually Long press UNIT button then press UNIT to select not Q2 Appear Error codes A2 Make sure if Hold appears on LCD ERROR CODES Code Description The electrode were...

Page 14: ...A if the unit is found to be defective it will be repaired or replaced at no charge OMEGA s WARRANTY does not apply to defects resulting from any action of the purchaser including but not limited to m...

Reviews: