FEATURES 

1) Professional look design accurate portable meters with large LCD display, with BNC connector.
2) All electrodes with BNC connector compatible to any other electrodes.
3) Hold function, low battery icon indicator, automatic power off in 15 minutes and may be disabled.
4) Built-in different temperature compensation selectable: Thermistor 30K, 10K ohm.
5) RFS (Recover to Factory Setting) function is included.
6) Conductivity calibration points: 1413uS. Calibration solution is included.
7) Conductivity electrode may convert to EC, TDS and Salinity concentration, Adjustable conductivity to TDS

conversion factor from 0.4 to 1.0. Three in one conductivity electrode (Cond/EC. TDS. Salinity) measurement.

8) Conductivity cell constant (K) selectable: 0.1, 1.0, 10 (Selectable and adjustable).
9) SD card capacity: 1GB to 8GB.
10) Tripod receptacle mountable design for long time monitoring purposes.
11) All sensors are well calibrated before shipment.
12) Warranty period of meter is two years. Electrode is one year.

SUPPLIED 

Meter

Electrode x 1pc (Cond/EC/TDS/Salt)

Battery-AAA x 3 pcs

1413uS buffer solution x1

9V adaptor

Black carrying case

8G SD card

SPECIFICATION 
Model 

CDH-SD11

Parameter 

Cond/EC/TDS /Salt 

Data logger sampling 
time 
Setting range 

Auto 

2 secs, 5 secs, 10 secs, 15 secs, 30 secs, 60 secs, 120 
secs, 300secs, 600 secs, 900 secs, 1800 secs, 1Hr 

Manual 

Press the ADJ button once will save data one time.@Set the 
sampling time to 0 second. 

Memory Card 

SD memory card 8G 

Data Hold 

Freeze the display reading. 

Meter dimension 

180mm x 58mm x 32mm (With BNC connector) 

Power supply 

AAA batteries x 3 pcs    /    9V AC/DC adaptor 

Temperature 

0~90

 

Temp. accuracy 

±0.5 

 

Conductivity

Measurement range 

0.05 uS/cm~1999 uS/cm (K=0.1) Option 

30 uS/cm~20 ms/cm (K=1) Supplied 

500 uS/cm ~199 mS/cm (K=10) Option 

Accuracy 

±1% of full scale + 1 digit 

Cond Calibration 

1 point    1413 uS calibration (K=1) 

1 point    12.8 mS calibration (K=10) 

Resolution 

0~19.99 uS/cm (0.01), 0~199.9 uS/cm (0.1), 0~1999 uS/cm (1) 

0~19.99 mS/cm (0.01), 0~199.9 mS/cm (0.1) 

Dimension 

12x120mm 

Electrode body 

ABS 

1

CDE-11G K=1 electrode replacement (Graphite type)

ADDITIONAL PURCHASE

CDE-11P K=10 option (Platinum type)

Summary of Contents for CDH-SD11

Page 1: ...e mail info omega com For latest product manuals www omegamanual info Shop online at omega com User sGuide CDH SD11 Conductivity TDS and Salinity Meter with Real Time SD Card Data Logger TM 2 e Fo...

Page 2: ...pecifications without notice Servicing North America U S A Omega Engineering Inc Headquarters Toll Free 1 800 826 6342 USA Canada only Customer Service 1 800 622 2378 USA Canada only Engineering Servi...

Page 3: ...two years Electrode is one year SUPPLIED Meter Electrode x 1pc Cond EC TDS Salt Battery AAA x 3 pcs 1413uS buffer solution x1 9V adaptor Black carrying case 8G SD card SPECIFICATION Model CDH SD11 Par...

Page 4: ...stant K value Temp coefficient compensation tds K factor value mode Long press again to escape the K value setting Recovery to factory setting under calibration mode SET CAL Conductivity Calibration H...

Page 5: ...uto power off function by pressing HOLD and PWR button n momentarily shows on screen now meter is at non sleep mode then turns to normal measurement Meter default auto power off 6 FACTORY RESET a Cond...

Page 6: ...r 2 Power on the meter by pressing PWR button 3 First to enter the cell constant K value of the Cond electrode is required e g Refer to the label on the top handle of the probe Long press SET UNIT to...

Page 7: ...right digit Long press SET button to finish calibration screen will show SA The adjustable range is 1 to 7 4 Or to escape setting by long press SET UNIT the bottom of screen will show ESC in few seco...

Page 8: ...measurement except to change the display unit from uS mS to PPM PPT Display will show TDS values at the same time the left bottom display will show the temp value of the measured solution NOTE If the...

Page 9: ...808 1837 1866 1896 1925 1954 1983 2013 2042 2071 2101 2103 CONDUCTIVITY STANDARD SOLUTION 12880 S cm at 25 C Conductivity reading will be affected by the temperature S cm S cm 5 10 15 16 17 18 19 20 2...

Page 10: ...c If the SD card is being used for the first time it is recommended that the card be formatted SD Card Formatting NOTE Always confirm that the device is compatible with the SD SDHC or SDXC memory card...

Page 11: ...essions the recorded data will be switched into the selected temperature unit 1 Power on the meter by AAA batteries or 9V AC DC adaptor 2 After inserting the SD card display will show icon SD on the b...

Page 12: ...for no data can be cleared or SD card isn t being inserted NOTE Once clear the data by long press CAL there is no way to recover the data If you want to keep the previous data rename the file MANUAL c...

Page 13: ...ensor type or adjust temperature manually Long press UNIT button then press UNIT to select not Q2 Appear Error codes A2 Make sure if Hold appears on LCD ERROR CODES Code Description The electrode were...

Page 14: ...A if the unit is found to be defective it will be repaired or replaced at no charge OMEGA s WARRANTY does not apply to defects resulting from any action of the purchaser including but not limited to m...

Reviews: