Factor= Actual TDS ÷ Actual Conductivity @ 25 

 

Solution 

Factor 

Water 

0.55~0.8 

Drinking water 

0.64 

Absolute pure water 

0.055 

Distilled water 

0.5 ~ 1.0 

To enter the factor by long press SET+UNIT, then short press MODE to toggle Con

t I F

tdS.   

  And go to select “tdS”and set the value.   

HOLD: 

to increase    ADJ: 

to decrease    SET: 

  to left digit    CAL: 

  to right digit 

Long press SET button to finish calibration, screen will show “SA”. The adjustable range is 0.05 to 1.00.   
Or to escape setting by long press SET+UNIT, the bottom of screen will show “ESC” few seconds and return 
to normal measurement mode. 

b) TDS measurement:
1) Under normal measurement mode, short press SET button to toggle Cond. measurement to TDS
measurement. 
2) The measuring procedures are same as above 2-1-c (page 7) conductivity measurement, except to change
the display 
  unit from uS, mS to PPM, PPT.   

Display will show TDS values at the same time the left bottom display will show the temp. value of the 
measured solution. 

NOTE: If the TDS value is ppm, the conductivity value must be in uS ; if the TDS value is in ppt, the 

conductivity value 

must be in mS. 

2-3 Salt 

a) Calibration

If the conductivity range already make

s

 the calibration completely (refer to 2-1-b at page 6 ) then the 

salt

 

measurement is not necessary to make the calibration again.

b) Salt measurement
1) Under normal measurement mode, short press SET button to toggle Cond. measurement to “salt”

measurement mode.

2) Hold the Cond. electrode handle by hand and let the sensing head Immersed wholly into the measured

solution. Shake the probe to let the electrode’s internal air bubble drift out from the sensing head.
Display will show salt values (ppt) at the same time the left bottom display will show the temp. value of the
measured solution.

6

Summary of Contents for CDH-SD11

Page 1: ...e mail info omega com For latest product manuals www omegamanual info Shop online at omega com User sGuide CDH SD11 Conductivity TDS and Salinity Meter with Real Time SD Card Data Logger TM 2 e Fo...

Page 2: ...pecifications without notice Servicing North America U S A Omega Engineering Inc Headquarters Toll Free 1 800 826 6342 USA Canada only Customer Service 1 800 622 2378 USA Canada only Engineering Servi...

Page 3: ...two years Electrode is one year SUPPLIED Meter Electrode x 1pc Cond EC TDS Salt Battery AAA x 3 pcs 1413uS buffer solution x1 9V adaptor Black carrying case 8G SD card SPECIFICATION Model CDH SD11 Par...

Page 4: ...stant K value Temp coefficient compensation tds K factor value mode Long press again to escape the K value setting Recovery to factory setting under calibration mode SET CAL Conductivity Calibration H...

Page 5: ...uto power off function by pressing HOLD and PWR button n momentarily shows on screen now meter is at non sleep mode then turns to normal measurement Meter default auto power off 6 FACTORY RESET a Cond...

Page 6: ...r 2 Power on the meter by pressing PWR button 3 First to enter the cell constant K value of the Cond electrode is required e g Refer to the label on the top handle of the probe Long press SET UNIT to...

Page 7: ...right digit Long press SET button to finish calibration screen will show SA The adjustable range is 1 to 7 4 Or to escape setting by long press SET UNIT the bottom of screen will show ESC in few seco...

Page 8: ...measurement except to change the display unit from uS mS to PPM PPT Display will show TDS values at the same time the left bottom display will show the temp value of the measured solution NOTE If the...

Page 9: ...808 1837 1866 1896 1925 1954 1983 2013 2042 2071 2101 2103 CONDUCTIVITY STANDARD SOLUTION 12880 S cm at 25 C Conductivity reading will be affected by the temperature S cm S cm 5 10 15 16 17 18 19 20 2...

Page 10: ...c If the SD card is being used for the first time it is recommended that the card be formatted SD Card Formatting NOTE Always confirm that the device is compatible with the SD SDHC or SDXC memory card...

Page 11: ...essions the recorded data will be switched into the selected temperature unit 1 Power on the meter by AAA batteries or 9V AC DC adaptor 2 After inserting the SD card display will show icon SD on the b...

Page 12: ...for no data can be cleared or SD card isn t being inserted NOTE Once clear the data by long press CAL there is no way to recover the data If you want to keep the previous data rename the file MANUAL c...

Page 13: ...ensor type or adjust temperature manually Long press UNIT button then press UNIT to select not Q2 Appear Error codes A2 Make sure if Hold appears on LCD ERROR CODES Code Description The electrode were...

Page 14: ...A if the unit is found to be defective it will be repaired or replaced at no charge OMEGA s WARRANTY does not apply to defects resulting from any action of the purchaser including but not limited to m...

Reviews: