30
c. Get a single reading of:
weight, A/D or temperature:
.
Get weight (after Filter2, not rounded to resolution).
?
Get weight (after Filter2, rounded to resolution).
>
Get A/D reading after Filter1.
<
Get A/D reading after Filter2.
T
Get temperature
d. Analog output: Read & write voltage:
@
Gets the analog output voltage (in Volts).
v
x
<c/r>
(lower case v)
Sets the analog output voltage to
x
Volts.
The analog output voltage is m
easured at pin 12 of CONN6
with respect to pin 13 which is ground.
e. Digital Outputs: Write outputs:
a
Turn on Output1 if Manual
A
Turn off Output1 if Manual
b
Turn on Output2 if Manual
B
Turn off Output2 if Manual
c
Turn on Output3 if Manual
C
Turn off Output3 if Manual
d
Turn on Output4 if Manual
D
Turn off Output4 if Manual
f. Digital Outputs: Read outputs:
O
(upper case o)
Card returns a string of the form ‘
xxxx
<c/r>’, where
x
is
either “1” or “0” representing the status of the 4 output
opto relays OUT4,OUT3,OUT2,OUT1 respectively.
g. Digital Inputs: Read inputs & toggling counter:
I
(upper case i)
Card returns a string of the form '
xxxx
<c/r>’, where
x
is
either “1” or “0” representing the digital input status of
IN4,IN3,IN2,IN1 respectively.
i
Gets a 16 bit ‘toggling counter’. The ‘toggling counter’
increments each time when input #2 is toggled
.