32
i. Misc.:
z
(lower case z)
Manually zero the gross weight. The effect of this function
is temporary — it expires upon card reset.
Response (versions 1.12, 3.09, 6.01, 7.00 and up):
‘z’<c/r>.
Z
(Upper case Z)
(versions 1.12, 3.09,
6.01, 7.00 and up)
Cancel the manual zero operation (the lower case ‘z’).
That is, return to the original calibration zero.
Response: ‘Z’<c/r>.
S
(upper case S)
System reset (software reset to perform a restart as if
turned off/on).
Response (versions 1.12, 3.09, 6.01, 7.00 and up):
‘S’<c/r> - possibly followed by additional binary
character(s).
Wait 6 seconds before accessing the board again.
Issue this command after changing any of the following
parameters:
* Auto transmit interval (parameter #20).
* Filters & Decimators (parameters #23-25).
* General Setpoints (parameters #101-104).
* Output modes (parameters #111-114).
F
(upper case F)
Start Fill-mode.
V
(upper case V)
Get version no. of DSP software.
Board’s response: “
LCIC-WIM/
V#.##”, where
#.## = Version Number.