www.microscopenet.com

 

 

17 

Note

 

  When using the 40X objective, you won

’t need to apply oil drop on the oil condenser. 

  When using the 100X objective, the condenser won

’t work well if no oil drop applied 

on the condenser. 

  When you apply the immersion oil with the 100X objective, do not let the immersion 

oil to contact with the dry objective lenses (especially the 40X). If the immersion oil 
is on the dry objectives lens, please use the lens cleaning paper to clean the 
objectives lens in time. The oil will damage the dry objective lenses. 

  After observing the specimen with the 100X objective, clean the 100X objective lens 

first. 

 
 
 
 

8.

 

Phase Contrast Kit Installation and Operation 
Instructions 

8.1  Phase contrast kit of A1PHIP: 

1)  Mounting the phase contrast objectives 

a. Take off all the objectives from the nosepiece. 
b. Install the phase contrast objectives onto the nosepiece following the steps in 

2.4

2)  Mounting the phase contrast condenser/annular ring disk 

a. Rotate the nosepiece and set the 10X objective in position.

 

b. Turn the coarse focus knob to raise the mechanical stage to the highest position 

without contacting the 10X objective.

 

c.  Turn the condenser control knob to lower the condenser to the lowest position.

 

d. Loosen the condenser lock thumb screw and remove the brightfield condenser as 

shown in 

Fig. 15 (a)

.

 

e. Insert the condenser/annular ring disk into the condenser holder, and tighten the 

thumb screw as shown in 

Fig. 15 (c)

.

 

f.  Re-adjust the condenser height to its normal position.

 

g. Re-adjust the mechanical stage height to its normal position.

 

 

 

 

 

Condenser Holder 

 

Fig.15 

Condenser Lock Thumb Screw 

 

Condenser Centering Screws 

 

(a) 

(b) 

(c) 

Summary of Contents for M8333S

Page 1: ...User Manual Trinocular Infinity Compound LED Microscope Model M8333S series MicroscopeNet com...

Page 2: ...1 Components Illustration 3 2 Installation 5 3 Operation 7 4 Specifications 10 5 Optional Parts 11 6 Troubleshooting Guide 13 7 Darkfield Condenser Installation and Operation Instructions 15 8 Phase...

Page 3: ...is located on a smooth level and firm surface 4 If any specimen solutions or other liquids splash onto the stage objective or any other component disconnect the power cord immediately and wipe up the...

Page 4: ...ng an excessive amount of solvents as this may cause problems with optical coatings or cemented optics or the flowing solvent may pick up grease making cleaning more difficult Oil immersion objectives...

Page 5: ...www microscopenet com 3 1 Components Illustration 11 12 13 14 15 16 17 7 18 7 1 2 3 4 5 6 7 8 9 10 19 7...

Page 6: ...4 Nosepiece 14 Objectives 24 Condenser 5 Slide Holder 15 Microscope Body 25 Aperture Diaphragm Ring 6 Mechanical Stage 16 Brightness Intensity Dial 26 Condenser Centering Screws 7 Light Collector 17...

Page 7: ...sure the viewing head is installed securely 2 2 Installation of the photo tube 1 Remove the cap on the top of the viewing head 2 Thread the photo tube into the top of the viewing head 2 3 Installation...

Page 8: ...The color filter is simply put into the filter holder 2 6 Connecting the power cord 1 Turn the power switch to the off position 2 Connect the power cord to the socket at the back of microscope body an...

Page 9: ...e center of viewing field Caution Be sure not to allow an objective to touch a specimen slide when changing objectives 3 3 Focusing 1 With the 10X objective in position raise the mechanical stage usin...

Page 10: ...3 5 Adjusting eyepiece diopter 1 Using the 10X objective and your right eye only observe your specimen through the eyepiece and bring it into focus by adjusting the focus knobs 2 Then observe the spe...

Page 11: ...ive clean the 100X objective lens first 3 7 Adjusting iris aperture diaphragm Turn the aperture diaphragm ring to adjust the aperture size 3 8 Adjusting focus knob tension The tightness of the focus k...

Page 12: ...e Double layer mechanical stage Dimension 5 1 2 x 4 1 4 140mm x 110mm Translation range 3 x 1 13 16 78mm x 46mm Condenser NA 0 9 0 25 with a flip up top lens built in aperture iris diaphragm Center ad...

Page 13: ...ion Operating System Software A35100U CMOS USB2 0 3584 x 2748 10 MP MS Windows Mac OS 10 8 and up Linux 2 6 or above Included A35140U 4096 x 3288 14 0MP A3550U3 CMOS USB3 0 2560 x 1922 5 0MP A3580U3 3...

Page 14: ...ntering Telescope A1PHIP Plan achromatic 10X with built in phase plate Plan achromatic 20X with built in phase plate Plan achromatic 40X with built in phase plate spring Plan achromatic 100X with buil...

Page 15: ...en Clean the lens with a lens cleaning paper Poor image quality or not able to get focused image No slide cover attached to the slide Attach a 0 17mm slide cover Slide cover is too thick or thin Use a...

Page 16: ...piece is not in the click stop position Revolve the nosepiece to the click stop position Lamp intensity is too high or low Adjust the light intensity by rotating the intensity control dial Slippage of...

Page 17: ...condenser till a dark spot showed in the viewing field as shown in Fig 12 a Turn the condenser translational centering screws to move the dark spot to the center as shown in Fig 12 b 3 Place the slide...

Page 18: ...owing in the viewing field Raise the condenser until the darkfield image showing in the viewing field 4 When using the 100X objective to observe the specimen Raise the condenser until the top is close...

Page 19: ...Mounting the phase contrast objectives a Take off all the objectives from the nosepiece b Install the phase contrast objectives onto the nosepiece following the steps in 2 4 2 Mounting the phase cont...

Page 20: ...19 to move the dark spot to the center until two ring images are coincided or Turn the adjustment knobs Fig 19 to move the dark spot to the center until two ring images are coincided f Remove the cent...

Reviews: