Olympus VISERA ENF-V2 Instructions Manual Download Page 56

52

Chapter 7  Cleaning, Disinfection and Sterilization Procedures for Endoscopic Instruments

VISERA RHINO-LARYNGO VIDEOSCOPE ENF-V2

• Do not store disinfected instruments. Use them immediately 

after disinfection or forward them to the following sterilization 
process. If stored instruments are used, refer to your local 
regulations and/or guidelines.

• Do not immerse instruments for more than 60 minutes in any 

liquids. If instruments are left immersed in liquids for a longer 
period, too much humidity inside of the endoscope will arise, 
which will cause malfunction.

When the regulation/guideline in your region requires 
disinfection prior to sterilization for the personnel in charge of 
reprocessing, perform the sterilization after the procedure of 
Section 6.6, “Automatic cleaning/disinfection” on page 41 or 
Section 7.7, “Disinfection” on page 63.

7.3

Precleaning

Preclean the endoscope immediately after the procedure. The following steps 
are to be performed when the light source and the video system center are 
turned OFF.

Endoscopic instrumentation must be meticulously cleaned 
prior to disinfection or sterilization. Thorough cleaning 
removes both micro-organisms and organic material. Failure 
to remove organic material decreases effectiveness of the 
disinfection or sterilization procedure. If the endoscope is not 
immediately precleaned, residual organic debris will dry on 
the instrument and effectively reprocessing will be difficult.

Equipment needed

Prepare the following equipment and wear appropriate personal protective 
equipment.

• Personal protective equipment
• Clean, lint-free cloth
• Detergent solution
• 500 cm

3

 (500 ml) containers

• De-ionized water (Aq. Dest.)

Summary of Contents for VISERA ENF-V2

Page 1: ...For PAL type only INSTRUCTIONS VISERA RHINO LARYNGO VIDEOSCOPE OLYMPUS ENF TYPE V2...

Page 2: ......

Page 3: ...sappeared or frozen endoscopic image 10 Examples of inappropriate handling 11 Chapter 1 Checking the Package Contents 13 Chapter 2 Instrument Nomenclature and Specifications 14 2 1 Nomenclature 14 2 2...

Page 4: ...nt 47 7 2 Cleaning disinfection and sterilization procedures 51 7 3 Precleaning 52 7 4 Transportation of the endoscope 54 7 5 Leakage testing 55 7 6 Manual cleaning 60 7 7 Disinfection 63 7 8 Rinsing...

Page 5: ...ymbol s shown on the package with the components the back cover of this instruction manual and or this instrument are as follows Refer to instructions Caution Type BF applied part Endoscope Manufactur...

Page 6: ...tent and probability Perform endoscopy and endoscopic treatment only when its potential benefits are greater than its risks Fully explain to the patient the potential benefits and risks of the endosco...

Page 7: ...to confirm that this instrument is compatible with the ancillary equipment being used Using incompatible equipment can result in patient and operator injury and or equipment damage This instrument com...

Page 8: ...de on page 73 If the irregularity is still suspected after inspection contact Olympus Prohibition of improper repair and modification This instrument does not contain any user serviceable parts Do not...

Page 9: ...ult Never operate the bending section perform suction insert or withdraw the endoscope s insertion tube without viewing the endoscopic image Patient injury can result Never operate the bending section...

Page 10: ...ed carefully If resistance to insertion is felt or the patient reports pain stop insertion immediately Otherwise operator and or patient injury can result or the endoscope could become lodged and be d...

Page 11: ...not illuminated unnecessarily If the endoscopic image becomes dimmer during the procedure it may indicate that blood or mucus is adhering to the light guide on the distal end of the endoscope Careful...

Page 12: ...rom the output socket of the light source and the endoscopic image will not be visible Do not pull the video cable during an examination The endoscopic image may not be visible anymore Do not coil the...

Page 13: ...and or may cause water leaks Do not attempt to bend the endoscope s insertion tube with excessive force Otherwise the insertion tube may be damaged To check the electromagnetic influence from other eq...

Page 14: ...e faulty contact can result Do not bend hit or twist the insertion section control section universal cord video connector and light guide connector The endoscope may be damaged and water leaks and or...

Page 15: ...d endoscopic treatment can be ensured through appropriate handling by the physician and the medical facility Examples of inappropriate handling are given below Inserting or withdrawing the endoscope a...

Page 16: ...12 Important Information Please Read Before Use VISERA RHINO LARYNGO VIDEOSCOPE ENF V2...

Page 17: ...onent is missing or you have any questions do not use the instrument immediately contact Olympus This instrument was not cleaned disinfected or sterilized before shipment Before using this instrument...

Page 18: ...RA RHINO LARYNGO VIDEOSCOPE ENF V2 Chapter 2 Instrument Nomenclature and Specifications 2 1 Nomenclature 4 Light guide connector 7 Venting connector Light guide Serial number Video cable Universal cor...

Page 19: ...VISERA RHINO LARYNGO VIDEOSCOPE ENF V2 1 Bending section 2 UP DOWN angulation control lever 3 Remote switches 3 Remote switches Control section Boot Insertion tube working length Distal end 9 Identif...

Page 20: ...t of the light source and transmits light from the light source to the endoscope 5 Video connector The video connector connects the endoscope to the output socket of the video system center and makes...

Page 21: ...idity 10 95 Atmospheric pressure 700 1060 hPa 0 7 1 1 kgf cm2 10 2 15 4 psia Transportation environment conditions during transportation and short term storage Ambient temperature 47 to 70 C 52 6 to 1...

Page 22: ...on Group 1 Class B This instrument complies with the EMC standard for medical electrical equipment edition 4 IEC 60601 1 2 2014 When connecting to an instrument that complies with a previous edition o...

Page 23: ...shooting If this instrument malfunctions do not use it Return it to Olympus for repair as described in Section 10 3 Returning the endoscope for repair on page 75 Using an endoscope that is not functio...

Page 24: ...the System chart in the Appendix and personal protective equipment such as eye wear face mask moisture resistant clothing and chemical resistant gloves before each use Refer to the respective instruc...

Page 25: ...nd light guide connector section for excessive scratching deformation or other irregularities 2 Inspect the boot and the insertion tube near the boot for bends twists or other irregularities 3 Inspect...

Page 26: ...t a point 10 cm from the distal end Push and pull gently to confirm that the angle between the bending section and the distal end is not loose 8 Inspect the objective lens at the distal end of the end...

Page 27: ...ystem center and display monitor as described in their respective instruction manuals Connection of the endoscope and ancillary equipment If the video connector and video system center are not connect...

Page 28: ...ly reprocessing agents 1 are aggressive and might effect the distal end of the endoscope When the effect progresses fogging of the inner lenses might appear which result in a foggy video image Any fur...

Page 29: ...endoscopic image is free from noise blur fog or other irregularities 5 Angulate the bending section and confirm that the endoscopic image is free from momentary disappearing or other irregularities In...

Page 30: ...system center OFF then without touching the angulation control lever slowly withdraw the endoscope from the patient If the angulation control mechanism or any other part of the system is not functioni...

Page 31: ...image on the video monitor disappears or freezes unexpectedly If the endoscopic image on the video monitor appears blurry of foggy unexpectedly If the angulation control mechanism is not functioning p...

Page 32: ...rikes the distal end Be sure to contact Olympus for repair 4 1 Insertion Holding and manipulating the endoscope The control section of the endoscope is designed to be held in the left hand The UP DOWN...

Page 33: ...an result or the endoscope could become lodged and be difficult to withdraw Transnasal insertion of the endoscope should be performed carefully If resistance to insertion is felt or the patient report...

Page 34: ...ndoscopic image Refer to the light source s instruction manual for instructions to adjust the brightness Observation of the NBI image All mucosal areas are to be viewed using traditional white light N...

Page 35: ...chemicals For the types and conditions of the means of cleaning disinfection and sterilization to be adopted please make judgments from your professional viewpoints Refer to the American National Stan...

Page 36: ...ope reprocessor follow their instruction manual Olympus does not assure the effect of cleaning disinfection and sterilization in endoscope reprocessors Contact the manufacturers of those disinfection...

Page 37: ...adequately ventilated Adequate ventilation protects against the buildup of toxic chemical fumes Store alcohol in an air tight container Alcohol stored in an open container may cause fires and will los...

Page 38: ...each method please contact Olympus If cleaning disinfection and sterilization methods not stated in this instruction manual are performed Olympus cannot guarantee the effectiveness safety and durabil...

Page 39: ...of material durability it means that the method can be adopted as a repetitive reprocessing procedure Just because the material durability is confirmed it does not mean a certain degree of microbiolo...

Page 40: ...lists those cleaning disinfection and sterilization procedures and agents that have been thoroughly tested on components of endoscopes and their accessories Methods validated in terms of biological ef...

Page 41: ...hout an identification mark for STERRAD 200 NX material compatibility on the control section is not compatible with STERRAD 200 NX When STERRAD 200 NX is used instrument damage and patient injury may...

Page 42: ...ral pH detergent or enzymatic detergent and follow the manufacturer s dilution and temperature recommendations Do not reuse detergent solutions Excessive detergent foaming can prevent fluid from adequ...

Page 43: ...be cleaned at room temperature Vapor generated by the ultrasonic cleaning device may contain infectious agents such as aerosol Always use a face screen and air exhaust during ultrasonic cleaning Exces...

Page 44: ...endoscopic instruments is certified by manufacturers and are approved by a competent authority Incompatible disinfectant solutions may considerably damage Olympus endoscopes and accessories For furth...

Page 45: ...l instruments have been securely fixed to the baskets Make sure that instruments do not touch each other Instruments with lumens must be attached to special tubes with rinsing devices Make sure that a...

Page 46: ...ment in the sterilization device Please verify the sterilization effects by using biological or chemical indicators Also follow the guidelines for sterilization issued by medical administrative author...

Page 47: ...re 20 EtO 80 CO2 Process Parameters Ethylene oxide gas sterilization Temperature 57 C Pressure 0 1 0 17 MPa Humidity 55 Exposure time 1 75 hours Ethylene oxide gas concentration 0 6 0 7 mg cm3 600 700...

Page 48: ...ave equipment let it cool down to room temperature Otherwise it may cause burns to the people handling the instrument Effective sterilization will not be possible if items are packed tightly together...

Page 49: ...on in a prevacuum sterilizer This method has been validated for its germicidal effectiveness with most of the Olympus instruments Olympus recommends that the instruments are autoclaved for 5 minutes a...

Page 50: ...kg 10 mg l SiO2 0 1 mg kg 1 mg l Fe 0 1 mg kg 0 2 mg l Cd 0 005 mg kg 0 005 mg l Pb 0 05 mg kg 0 05 mg l Traces of heavy metals except Fe Cd Pb 0 1 mg kg 0 1 mg l Cl 0 1 mg kg 2 mg l P2O5 0 1 mg kg 0...

Page 51: ...cleaning disinfection and sterilization procedures for the endoscopic instruments that are related to the safety and health of patients Regarding the maintenance procedure of equipments for reprocessi...

Page 52: ...500 cm3 500 ml containers Detergent solution Clean lint free cloths Large non metallic basins with a tight fitting lid for detergent and disinfectant solution Disinfectant solution Sterile lint free...

Page 53: ...6 When performing gas sterilization ethylene oxide gas STERRAD 50 100S etc the ETO cap must be attached to the venting connector to equalise the outer and inner pressure see Figure 7 2 Figure 7 2 Leak...

Page 54: ...are subject to wear Should any irregularity be observed use a new one Using defective equipment may result in ineffectual reprocessing and cause equipment damage Inspection of the ETO cap Confirm tha...

Page 55: ...e endoscope Section 7 4 on page 54 Withdrawal of the endoscope Precleaning Section 7 3 on page 52 Leakage testing Section 7 5 on page 55 Manual cleaning Section 7 6 on page 60 Rinsing and drying Autom...

Page 56: ...the procedure of Section 6 6 Automatic cleaning disinfection on page 41 or Section 7 7 Disinfection on page 63 7 3 Precleaning Preclean the endoscope immediately after the procedure The following step...

Page 57: ...stretch or severely damage the insertion tube and the covering of the bending section Gently wipe off debris on the insertion tube with a clean lint free cloth soaked in detergent solution Wipe from...

Page 58: ...e insertion tube securely but gently without squeezing with the video connector in the other hand see Figure 7 6 Figure 7 6 Transport reusable products except endoscope from the point of use to the re...

Page 59: ...d wear appropriate personal protective equipment Leakage testing when using MB 155 prepare the following equipment Never connect or disconnect the leakage tester s connector cap while immersed Doing s...

Page 60: ...emaining on either component could cause endoscope malfunction When the leakage tester connector WA23070A MB 155 is in place the covering of the bending section will expand as the air pressure inside...

Page 61: ...ale area press the pressure release lever to let the air escape Otherwise equipment damage may result 1 Fill a basin with clean water Use a basin which is at least 25 W 10 H 50 D cm in size and deep e...

Page 62: ...us water leakage or the leakage tester may be damaged Stop leakage testing immediately Water may enter the endoscope with no pressure inside if it is still immersed in water As a result problems other...

Page 63: ...the leakage tester s connector cap 4 Attach the leakage tester to the venting connector on the light guide connector see Figure 7 9 5 With the leakage tester connected immerse the endoscope in water a...

Page 64: ...sive force when cleaning the endoscope Regarding the detergent solution refer to Section 6 2 Detergent solution on page 38 After completing the leakage test perform manual cleaning according to the pr...

Page 65: ...ternal surface 1 Immerse the endoscope in the detergent solution 2 With the endoscope immersed use a soft brush or lint free cloth to thoroughly brush and wipe all external surfaces of the endoscope C...

Page 66: ...t free cloth gently wipe off and dry the external surfaces of the endoscope 2 Inspect the endoscope for residual debris Should debris remain repeat the procedures given in this Section Presoak for exc...

Page 67: ...ge 63 or Section 7 10 Sterilization on page 67 7 7 Disinfection All disinfection steps should be performed with the endoscope completely immersed Otherwise disinfectant solution may not adequately con...

Page 68: ...l disinfection steps should be performed with the endoscope and all equipment completely immersed If the equipment is not completely immersed disinfectant solution may not adequately contact all surfa...

Page 69: ...d water Aq Dest Do not use tap water for rinsing because it might be chlorinated Regarding the rinsing water quality refer to Section 6 5 Rinsing water on page 40 After disinfection rinse the endoscop...

Page 70: ...mmerse the endoscope in de ionized water Aq Dest Using a clean lint free cloth thoroughly rinse and wipe all outer surfaces 2 Remove the endoscope from water 3 Use a sterile lint free cloth moistened...

Page 71: ...covering of the bending section Regarding the condition of ethylene oxide gas sterilization refer to Section 6 7 Ethylene oxide gas sterilization on page 42 1 Dry the endoscope before ethylene oxide...

Page 72: ...0 steam sterilize autoclave according to the procedures given below 1 Seal the individual parts or equipment separately in packages appropriate for steam sterilization autoclave according to your hosp...

Page 73: ...RHINO LARYNGO VIDEOSCOPE ENF V2 Chapter 8 Maintenance Procedures of Equipments for Reprocessing Clean and decontaminate the equipment after use If necessary perform ultrasonic cleaning For compatible...

Page 74: ...g case Use the carrying case only for shipping the endoscope Routinely storing the endoscope in a humid non ventilated environment such as the carrying case may present an infection control risk 1 Bef...

Page 75: ...the carrying case again before use If the endoscope is not cleaned disinfected or sterilized the patient might be infected The carrying case cannot be cleaned disinfected or sterilized Clean and disin...

Page 76: ...described remedial action stop using the endoscope and send it to Olympus for repair Olympus does not repair accessory parts If an accessory part becomes damaged contact Olympus to purchase a replacem...

Page 77: ...3 Returning the endoscope for repair on page 75 Image quality or brightness Other Irregularity description Possible cause Solution There is no video image Not all the power switches are ON Turn ON all...

Page 78: ...eeding and or perforation When the endoscopic image appears on the monitor 1 Turn OFF all equipment except the video system center light source and video monitor 2 Carefully withdraw the endoscope whi...

Page 79: ...l or at Olympus Olympus is not liable for any injury or damage which occurs as a result of repairs attempted by non Olympus personnel Before returning the endoscope for repair contact Olympus With the...

Page 80: ...may also be compatible for use in combination with this instrument For further details contact Olympus If combinations of equipment other than those shown below are used the full responsibility is as...

Page 81: ...center OTV S7V Video system OTV SI Endoscope ENF V2 Video monitor OEV141 OEV142 OEV143 OEV201 OEV202 OEV203 OEV181H OEV191 EVIS EXERA II xenon light source CLV 180 EVIS EXERA II video system center C...

Page 82: ...eaning and disinfection equipment Maintenance unit MU 1 ETO cap MB 156 Leakage tester MB 155 Leakage tester WA23070A Ultrasonic cleaner KS 2 Sterilization tray WA05991A ETD series 1 1 For any other de...

Page 83: ...y for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment Radiated emissions CISPR 11 Class B This instrument s RF...

Page 84: ...2 4 6 kV Air 2 4 8 kV Same as left Floors should be made of wood concrete or ceramic tile that hardly produces static If floors are covered with synthetic material that tends to produce static the rel...

Page 85: ...UT is the a c mains voltage prior to application of the test level Power frequency 50 60 Hz magnetic field IEC 61000 4 8 30 A m 50 Hz 60 Hz 3 A m 50 Hz 60 Hz Same as left It is recommended to use this...

Page 86: ...ation distance in meters m Proximity magnetic field from RF communication equipment IEC 61000 4 3 Refer to the table of the next page Same as left a Field strength from fixed transmitters such as base...

Page 87: ...sed for EMC compliance testing Refer to the instruction manuals for each piece of equipment Test frequency MHz Band MHz Modulation 1 Maximum power W Immunity test level V m 385 380 390 Pulse modulatio...

Page 88: ......

Page 89: ...reproduced or distributed without the express written permission of OLYMPUS MEDICAL SYSTEMS CORP OLYMPUS is a registered trademark of OLYMPUS CORPORATION Trademarks product names logos or trade names...

Page 90: ...Valley Road 12 01 04 Valley Point Office Tower Singapore 248373 Fax 6834 2438 Telephone 6834 0010 A8F Ping An International Financial Center No 1 3 Xinyuan South Road Chaoyang District Beijing 100027...

Reviews: