Chapter 5 Reprocessing: General Policy
VISERA RHINO-LARYNGO VIDEOSCOPE ENF-V2
• This instrument is not durable, or does not have sufficient
durability against the respective methods stated in the
guidelines of each country for destroying or inactivating
prions. For information on the durability against each
method, please contact Olympus. If cleaning, disinfection
and sterilization methods not stated in this instruction manual
are performed, Olympus cannot guarantee the effectiveness,
safety and durability of this instrument. Make sure to confirm
that there is no abnormality before use, and use under
responsibility of a physician. Do not use if any abnormality is
found.
Before immersing the endoscope in cleaning or disinfectant
solution, confirm that the ETO cap is not attached to the
venting connector.