Chapter 7 Cleaning, Disinfection and Sterilization Procedures for Endoscopic Instruments
VISERA RHINO-LARYNGO VIDEOSCOPE ENF-V2
Inspection of the equipment
For inspection of other equipment than that mentioned below, refer to the
instruction manual for the equipment being used.
All equipment mentioned below are items that are subject to
wear. Should any irregularity be observed, use a new one.
Using defective equipment may result in ineffectual
reprocessing, and cause equipment damage.
Inspection of the ETO cap
Confirm that the ETO cap is free from scratches, flaws and debris (see Figure
7.2).
Inspection of the leakage tester (WA23070A)
1.
Check for cracks, scratches, flaws, debris and other damage (see Figure
7.3).
2.
Press the hand pump and confirm that the leakage tester is emitting air from
the adapter (see Figure 7.4).
Figure 7.4
Adapter
Air
Hand pump