84
Appendix
ELECTROSURGICAL UNIT PSD-30
Item
Specification
Classification
(electromedical
equipment)
Type of
protection
against electric
shock
Class
I
(3 pin power cord)
Degree of
protection
against electric
shock
TYPE CF (the patient leakage current below
10 µA)
Note: Direct application to the heart should not
be attempted as spark discharge during
electrosurgery may affect the cardiac function
of the patient (refer to dangers, warnings and
cautions and EN standard (EN60601-1)).
Degree of
protection
against
explosion
Never use the electrosurgical unit where there
is a risk of flammable gases.
Additional
Defibrillation-protected equipment.
Output circuit
Neutral electrode isolated from earth at high
frequency.
Power supply
Voltage
220, 230, 240 V
Frequency
50/60 Hz
Input current
2 A
Voltage
fluctuation
Within
±
10%
Fuse rating
3.15 A, 250 V
Fuse size
5.0 × 20 mm
Size
Dimensions
295 mm (W) × 160 mm (H) × 420 mm (L)
Weight
7.8 kg