Chapter 4 Inspection
37
ELECTROSURGICAL UNIT PSD-30
Figure 4.3
Figure 4.4
4.
Inspect both fuses.
Use only the fuses listed below. Otherwise, fire or equipment
damage can result.
5.
Even if only one fuse has blown, always replace both of them (see Figure
4.4).
If both fuses have not blown, contact Olympus.
6.
Insert the fuse box into the electrosurgical until it clicks into position.
7.
Reconnect the power cord and press the power switch. Confirm that the
power switch is lit.
Olympus
MAJ-1052
Littel
2183.15
Fuse
Fuse box
Fuse replacement