Chapter 3 Installation and Connection
ELECTROSURGICAL UNIT PSD-30
3.5
Connection of the S-cord (when using an
endoscope with a S-cord connector mount)
Connect the S-cord (MAJ-813) firmly between the S-cord
connector on the front panel and the S-cord connector mount
of the endoscope. Otherwise, the high frequency may not be
output during treatment and, in this case, mechanical tissue
cutting by the electrosurgical accessory may occur and
cause bleeding or perforation of the patient.
1.
Confirm that the S-cord is free of scratches and cracks.
2.
With the arrow mark on the S-cord’s plug on the unit side facing upward,
insert the plug into the S-cord connector on the electrosurgical unit front
panel, until it clicks (see Figure 3.9).
Figure 3.9
3.
Connect the S-cord’s plug on the endoscope side to the S-cord connector
mount on the endoscope.
Figure 3.10
Arrow mark
Fastener ring
Endoscope
socket
Plug on the
endoscope side
S-cord connector
receptacle