4-2  Stage

Fig. 6

Fig. 7

1

    Placing the Specimen

(Figs. 6 & 7)

When using a mechanical stage

1. Set the stage plate @. (Fig. 6)

2. Place the specimen ² on the stage plate with the observation surface

facing downward. (Fig. 6)

# When setting the stage plate on the mechanical stage, use the stage

plate holder ³ provided with the mechanical stage.

# Do not use the   35 mm petri dish holder | and slide glass holder 

5

.

When using a plain stage

1. Select the stage center plate @ according to the specimen size and

place it on the hole at the center of the stage.

2. Place the specimen, with the surface to be observed facing down, gently

on the stage center plate.

With the GX-CP stage center plate, hold the specimen by applying opti-

mum weight with the specimen holder ².

# Do not forcefully press the specimen holder against the stage cen-

ter plate for it may deform the plate.

When the specimen is heavier than 1 kilogram, special care is re-

quired because the stage center plate could be deformed or the

operability of the stage feed knobs may deteriorate.

}If the stage center plate is removed, a specimen as heavy as about 5

kilograms can be placed on the stage. Set up the microscope on a

sturdy desk. In the environment susceptible to external vibration, it is rec-

ommended to use the portable anti-vibration table.

Fig. 8

2

    Moving the Specimen

(Fig. 8)

When using the plain stage, move the specimen directly by hand or

manually move the stage plate on which the specimen is placed to the

required position. When using the mechanical stage, turn the feed rings

on it to move the specimen.

# The maximum weight of a specimen that can be placed on the

stage plate @ when using the mechanical stage is 400 grams. Speci-

men weighing more than 400 grams will prevent the stage plate•fs

movement from following the rotation of the feed rings.

# Do not place a heavy object including a specimen on the area ²

directly above the feed rings of the mechanical stage. This may ad-

versely affect the mechanical stage and could cause the specimen

to drop.

@

²

@

²

@

²

3

4

5

3

Summary of Contents for GX41

Page 1: ...GX41 To ensure the safety obtain optimum performance and to familiarize yourself fully with the use of this microscope we recommend that you study this manual thoroughly before operating the micro sc...

Page 2: ...comply with the limits for a Class A digital device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference when the product is op...

Page 3: ...7 18 20 4 1 Microscope Frame 6 4 2 Stage 7 8 4 3 Observation Tube 9 12 4 4 Illumination Unit 13 14 19 10 Correct assembly and adjustments are indispensable for the microscope to manifest its full perf...

Page 4: ......

Page 5: ...llips 5761 Halogen bulb 6V 30WHAL Philips 5761 6 Always use the power cord provided by Olympus If no power cord is provided please select the power cord by referring to the section PROPER SELECTION OF...

Page 6: ...m 2 Many parts of the exterior are made of plastic Wipe the microscope with a clean cloth only Do not use organic solvents to clean non optical components If smudges are difficult to remove wipe them...

Page 7: ...re usable even when they are not mentioned below For these modules refer to the latest catalogues or contact Olympus Stage Center Plate GX CP IX2 GCP IX CP50 U CBI30 2 U BI30 2 U CTR30 2 U TR30 2 U CT...

Page 8: ...htness control knob Page 6 Coarse tension adjustment ring Page 6 1 Main switch Page 6 7 Coarse adjustment knob Page 6 5 Specimen holder Page 7 4 Stage plate Page 7 Stage plate holder a Aperture iris d...

Page 9: ...Adjust the interpupillary distance and diopter Engage the objective of the desired magnification in the light path and focus on the specimen Adjust the illumination brightness Engage the desired filte...

Page 10: ...ositions 3 Adjusting the Tension of the Coarse Adjustment Knob Fig 5 Be sure to use the coarse tension adjustment ring to adjust the tension of the control adjustment knob How to adjust the tension Tu...

Page 11: ...is re quired because the stage center plate could be deformed or the operability of the stage feed knobs may deteriorate If the stage center plate is removed a specimen as heavy as about 5 kilograms...

Page 12: ...e stage 1 Using the spanner provided with the observation position confirma tion mirror attach the mirror to either of the tapped holes on the left and right remote ends of the stage 2 Adjust the mirr...

Page 13: ...ex dots horizontal adjust so that the index dots come on the extension of one of the horizontality lines inscribed on the pivot When your interpupillary distance is other than 50 60 70 and 75 adjust s...

Page 14: ...ive in the right path and turn only the helicoids of the right eyepiece to focus on the specimen The precision can be improved by repeating steps 1 and 2 several times 3 Look through the left eyepiece...

Page 15: ...set Fig 17 However if the eyepiece does not offer the helicoidal adjustment facility observers with poor eyesight are recommended to wear eyeglasses be cause they cannot focus the eye on the micromet...

Page 16: ...I3 5 to 35 Never attempt to force the binocular section past the upper or lower stop position Applying excessive force could destroy the limiting mechanism Tilt adjustment is possible with specified e...

Page 17: ...dges with both index fingers and place it in an appropriate filter groove The filter grooves are positioned obliquely to prevent flares First remove the stage center plate and then place a required fi...

Page 18: ...tproof glass is still dirty wipe using a cotton swab moistened with a cleaning fluid To approach the dust glass lower the revolving nosepiece Fig 24 2 Using the Aperture Iris Diaphragm Fig 24 The aper...

Page 19: ...th high reflective surface and minimized polarization like a mirror on the stage plate 5 Slightly rotate the polarizer until the field of view is at its dimmest level so that the cross Nicol position...

Page 20: ...CCD 2 3 in CCD 1 2 in CCD When the 0 5X TV adapter is used When the 1X TV adapter is used The magnification of the TV adapter to be used is determined by the size of the CCD incorporated in the TV cam...

Page 21: ...ield of view The specimen is dirty Clean the specimen The eyepiece is dirty Clean the eyepiece d The image appears shiny The aperture iris diaphragm is stopped down Open the aperture iris diaphragm e...

Page 22: ...t appropriately 4 Observation Tube a The fields of view of the left and right eyepieces do not match The interpupillary distance is not correctly adjusted Adjust it correctly The diopter is not correc...

Page 23: ...ween binocular and straight tubes variable in 3 steps 100 0 20 80 0 100 Tilting tubes U CTBI Field number 18 tube inclination 30 60 interpupillary distance adjustment range 48 75 mm 10X eyepieces are...

Page 24: ...HARACTERISTICS Magnification 5X 10X 20X 50X 100X Numerical aperture 0 10 0 25 0 40 0 75 0 90 Working distance mm 20 0 10 6 1 3 0 38 0 21 Resolution m 3 36 1 34 0 84 0 45 0 37 WHB10X Field number 20 To...

Page 25: ...r AB4488 Microscope frame GX41F Objectives MPlan series Power cord Lamp socket cord Halogen bulb 6V30WHAL L 6V30WHAL Lamp socket U LS30 4 Light guide LG SF Light guide adapter SZX TLGAD Light source L...

Page 26: ...e After setting the main switch to OFF and unplugging the power cord from the wall outlet allow the old bulb and lamp socket to cool before replacing the bulb with a new one of the designated type Fig...

Page 27: ...stage is not necessary if you are going to observe a large object or if the mechanical stage might obstruct your view In this case use a stage center plate matching the specimen size as required Fig...

Page 28: ...cord or cable may melt and result in shock hazard To prevent this distribute the cords and cables apart from the lamp socket Fig 35 Fig 36 Fig 37 Intermediate attachment 8 Mounting the Intermediate A...

Page 29: ...ing Table 1 Certified Cord A power supply cord should be certified by one of the agencies listed in Table 1 or comprised of cordage marked with an agency marking per Table 1 or marked per Table 2 The...

Page 30: ...ter sterreichischer Verband f r Elektrotechnik VE Danmarks Elektriske Materialkontroll DEMKO National Standards Authority of Ireland NSAI Norges Elektriske Materiellkontroll NEMKO Asociacion Electrote...

Page 31: ...If you have any questions please contact Olympus Check results Date Check items 1 More than 8 years have passed since original purchase or the total power ON time exceeds 20 000 hours 2 Lamp does not...

Page 32: ...Bongeunsa ro Gangnam gu Seoul 06153 Korea Distributed by Manufactured by 48 Woerd Avenue Waltham MA 02453 U S A Shinjuku Monolith 2 3 1 Nishi Shinjuku Shinjuku ku Tokyo 163 0914 Japan www olympus glo...

Reviews: