10-003591-01EN [Q7780031], Rev. 4, January 2020
Chapter 2
24
2.2.1
Setup Characteristics
To create an instrument setup, make the following selections on your OmniScan
instrument:
•
FA1-type probe model
•
Depending on the application, one of the following three types of wedges:
— FlexoFORM scanner applications: SFA1-FLEXO
— Automated inspections: SFA1-AUTO
— Small diameter inspections: SFA1-SMALL
•
Select
Plate
in the
Specimen Type
parameter field (see Figure 2-1 on page 24).
Although the probe has a concave shape, it is important to select
Plate
, even if you
are inspecting a curved surface.
Figure 2-1 Selecting Plate in the Specimen Type field
2.2.2
Setting the Parameters for a Symmetrical Scan
For pipe elbow inspections, Olympus recommends the symmetrical scan pattern for a
full 360° inspection around the circumference (see Figure 2-2 on page 25). The
OmniScan instrument must be correctly set up to enable this type of inspection. For
details about alternative inspection patterns for straight pipe surfaces, see “Setting Up
to Inspect a Straight Pipe — Unidirectional or Bidirectional Scan” on page 75.
Summary of Contents for FlexoFORM
Page 6: ...Table of Contents vi 10 003591 01EN Q7780031 Rev 4 January 2020 ...
Page 8: ...10 003591 01EN Q7780031 Rev 4 January 2020 List of Abbreviations viii ...
Page 12: ...10 003591 01EN Q7780031 Rev 4 January 2020 Labels and Symbols 4 ...
Page 24: ...10 003591 01EN Q7780031 Rev 4 January 2020 Important Information Please Read Before Use 16 ...
Page 26: ...10 003591 01EN Q7780031 Rev 4 January 2020 Introduction 18 ...
Page 30: ...10 003591 01EN Q7780031 Rev 4 January 2020 Chapter 1 22 ...
Page 44: ...10 003591 01EN Q7780031 Rev 4 January 2020 Chapter 3 36 ...
Page 82: ...10 003591 01EN Q7780031 Rev 4 January 2020 Chapter 7 74 ...
Page 86: ...10 003591 01EN Q7780031 Rev 4 January 2020 Appendix B 78 ...
Page 90: ...10 003591 01EN Q7780031 Rev 4 January 2020 List of Tables 82 ...
Page 94: ...10 003591 01EN Q7780031 Rev 4 January 2020 Index 86 ...