10-003591-01EN [Q7780031], Rev. 4, January 2020
Chapter 5
52
Figure 5-4 Changing the probe’s o-ring seal on the wedge
To change the o-ring seal on the water connection
1.
Use an appropriate tool such as a small flathead screwdriver to carefully lift and
pull off the old o-ring seal without damaging the mounting surface (see
Figure 5-5 on page 52).
2.
Install the new o-ring seal.
Figure 5-5 Changing the o-ring seal for the water connection
5.7
Changing the Water Tube
The water tube on the scanner can be changed if it is damaged.
To change the water tube
1.
Remove the screw that fixes the water connector to the scanner (see Figure 5-6 on
page 53).
2.
Pull the connector off the water tube.
3.
Pull back the cable sleeve to access the tube, disconnect it from the union
connector, and pull it out of the scanner.
Summary of Contents for FlexoFORM
Page 6: ...Table of Contents vi 10 003591 01EN Q7780031 Rev 4 January 2020 ...
Page 8: ...10 003591 01EN Q7780031 Rev 4 January 2020 List of Abbreviations viii ...
Page 12: ...10 003591 01EN Q7780031 Rev 4 January 2020 Labels and Symbols 4 ...
Page 24: ...10 003591 01EN Q7780031 Rev 4 January 2020 Important Information Please Read Before Use 16 ...
Page 26: ...10 003591 01EN Q7780031 Rev 4 January 2020 Introduction 18 ...
Page 30: ...10 003591 01EN Q7780031 Rev 4 January 2020 Chapter 1 22 ...
Page 44: ...10 003591 01EN Q7780031 Rev 4 January 2020 Chapter 3 36 ...
Page 82: ...10 003591 01EN Q7780031 Rev 4 January 2020 Chapter 7 74 ...
Page 86: ...10 003591 01EN Q7780031 Rev 4 January 2020 Appendix B 78 ...
Page 90: ...10 003591 01EN Q7780031 Rev 4 January 2020 List of Tables 82 ...
Page 94: ...10 003591 01EN Q7780031 Rev 4 January 2020 Index 86 ...