TROUBLESHOOTING CLV-160
Lamp check
The light source xenon lamp is not lit.
Power supply
check
Panel operation
check
Does the audible
warning ring?
Yes
Check the temperature
switch.
• ON if normal.
No
No
Does the audible
warning ring?
Yes
Replace the UPCL16CR00.
(Replace the main board.)
Does the xenon
lamp light up?
No
Yes
Check the lamp mounting.
• The lamp is correctly mounted.
The heat compound is correctly
p a i n t e d .
Yes
Does the xenon
lamp light up?
No
Replace the lamp by a
brand-new one (Y1064S).
Yes
Yes
Does the xenon
lamp light up?
No
Check around the lamp
housing.
Does the xenon
lamp light up?
• Unnecessary things (metallic pieces, etc.)
• High voltage pulse leaks.
• Connection with the electrode F and
electrode R.
The round terminal is close to the lamp
h o u s i n g .
No
A
B
4-4