Replacing a Phased Array Probe
To replace a phased array ultrasonic probe, the probe assembly must be removed and
partially disassembled.
To replace a phased array probe
1.
Turn off the couplant flow.
2.
Disconnect the couplant tubes from the couplant source.
3.
Disconnect the probe cable from the instrument.
To prevent any damage to the Hypertronics connector pins, be sure to always install
the protective cover when the probe cable is disconnected.
4.
Remove the scanner from the inspection surface.
5.
Lift the swing arm until the latch engages into its higher position (see “To latch
the swing arm” on page 28).
CAUTION