DMTA048-01EN, Rev. C, February 2013
52 Chapter 4
CFU-PWZ: Remote
control’s light is not
on when the remote
control’s power
button is pressed
down and the
pump’s motor does
not turn on.
Bad electric
connection
Check these connections:
Power plug in AC power outlet
(wall)
•
power cable in the power
supply unit’s AC In
connection
•
power supply connector
in the remote control’s
DC
IN
receptacle
Table 4 Troubleshooting
(continued)
Problem Description
Possible Cause
Solution
Downloaded from
Summary of Contents for CFU-PWZ
Page 18: ...DMTA048 01EN Rev C February 2013 14 Introduction Downloaded from ManualsNet com search engine...
Page 26: ...DMTA048 01EN Rev C February 2013 22 Chapter 1 Downloaded from ManualsNet com search engine...
Page 42: ...DMTA048 01EN Rev C February 2013 38 Chapter 2 Downloaded from ManualsNet com search engine...
Page 46: ...DMTA048 01EN Rev C February 2013 42 Chapter 3 Downloaded from ManualsNet com search engine...
Page 64: ...DMTA048 01EN Rev C February 2013 60 Chapter 5 Downloaded from ManualsNet com search engine...
Page 72: ...DMTA048 01EN Rev C February 2013 68 Index Downloaded from ManualsNet com search engine...