DMTA048-01EN, Rev. C, February 2013
Maintenance and Troubleshooting 49
Motor is running but
no water goes into
the inlet tube (cannot
prime)
The inlet tube or the
outlet tube has a kink
Remove the kink.
Water outlet holes
(for the scanner or
wedge) are
obstructed
Remove dirt from the water outlet
holes.
Dirt on inlet filter
Clean the inlet filter by following
the procedure in section 4.1
“Maintaining the Water Inlet
Tube’s Suction Filter” on page 43.
Bad inlet tube
connection
Follow the connection procedure
in section 2.3.1 “Connecting the
Water Supply Tube” on page 29.
Inlet filter is above
the water line
Fill up the water tank if it is
empty or sink the inlet filter to the
bottom of the tank.
Dirt on
INLET
connector
Remove dirt from the connector.
CFU03 or CFU05:
Valve on the pump
partially or fully
closed
Turn the pump valve counter
clockwise to allow more water
flow.
CFU-PWZ: Valve on
the PipeWIZARD
partially or fully
closed
Turn the PipeWIZARD valve
clockwise to allow more water
flow.
Table 4 Troubleshooting
(continued)
Problem Description
Possible Cause
Solution
Downloaded from
Summary of Contents for CFU-PWZ
Page 18: ...DMTA048 01EN Rev C February 2013 14 Introduction Downloaded from ManualsNet com search engine...
Page 26: ...DMTA048 01EN Rev C February 2013 22 Chapter 1 Downloaded from ManualsNet com search engine...
Page 42: ...DMTA048 01EN Rev C February 2013 38 Chapter 2 Downloaded from ManualsNet com search engine...
Page 46: ...DMTA048 01EN Rev C February 2013 42 Chapter 3 Downloaded from ManualsNet com search engine...
Page 64: ...DMTA048 01EN Rev C February 2013 60 Chapter 5 Downloaded from ManualsNet com search engine...
Page 72: ...DMTA048 01EN Rev C February 2013 68 Index Downloaded from ManualsNet com search engine...