DMTA048-01EN, Rev. C, February 2013
Maintenance and Troubleshooting 51
Presence of air
bubbles in the outlet
tube
Dirt on
INLET
connector
Remove dirt from the connector.
Bad inlet tube
connection
Follow the connection procedure
in section 2.3.1 “Connecting the
Water Supply Tube” on page 29.
Damaged or
punctured water inlet
tube
Replace the water inlet tube.
Contact your local Olympus NDT
Service Center for a replacement.
The inlet tube has a
kink
Remove the kink.
Dirt on inlet filter
Clean the inlet filter by following
the procedure in section 4.1
“Maintaining the Water Inlet
Tube’s Suction Filter” on page 43.
Inlet filter is above
the water line
Fill up the water tank if it is
empty or sink the inlet filter to the
bottom of the tank.
Pump has just been
primed and air is
trapped in the tube
Fully open the flow control valve
and let the pump run for two
minutes.
CFU-PWZ: Remote
control’s light is not
on when the remote
control’s power
button is pressed
down and the pump
is functioning
normally
Burned light
Contact your local Olympus NDT
Service Center for repair.
Table 4 Troubleshooting
(continued)
Problem Description
Possible Cause
Solution
Downloaded from
Summary of Contents for CFU-PWZ
Page 18: ...DMTA048 01EN Rev C February 2013 14 Introduction Downloaded from ManualsNet com search engine...
Page 26: ...DMTA048 01EN Rev C February 2013 22 Chapter 1 Downloaded from ManualsNet com search engine...
Page 42: ...DMTA048 01EN Rev C February 2013 38 Chapter 2 Downloaded from ManualsNet com search engine...
Page 46: ...DMTA048 01EN Rev C February 2013 42 Chapter 3 Downloaded from ManualsNet com search engine...
Page 64: ...DMTA048 01EN Rev C February 2013 60 Chapter 5 Downloaded from ManualsNet com search engine...
Page 72: ...DMTA048 01EN Rev C February 2013 68 Index Downloaded from ManualsNet com search engine...