81
BX53M
Phenomena
Cause
Remedy
Page
Number of beeps
CB
LIM /
HS
MIX
None
/
Turns ON
The MIX slider for reflected light observation (U-MIXR) is not
engaged in the light path.
Engage the MIX slider for reflected light observation in the light
path.
If the POSITION indicator of the hand switch (BX3M-HS) keeps
lighting OUT (IN does not light.) even though the MIX slider for
reflected light observation is engaged in the light path, contact
Olympus.
66
None
/
Turns ON
The illumination of the MIX slider for reflected light
observation (U-MIXR) is turned OFF.
Or the hand switch (BX3M-HS) is damaged.
Press the ON button of the hand switch (BX3M-HS).
If the illumination does not turn ON even though the ON button
is pressed, contact Olympus.
66
None
/
Turns ON
The light intensity value of the MIX slider for reflected light
observation (U-MIXR) is 0%.
Or the hand switch (BX3M-HS) is damaged.
Press the light intensity button (
) of the hand switch (BX3M-
HS).
If the illumination does not turn ON even though the light
intensity button is pressed, contact Olympus.
67
g) The CUBE indicator or the OB indicator does not light even though following operations were performed.
· Select the observation method using the coded reflected light illuminator.
· Select the objective by turning the motorized nosepiece or the coded nosepiece.
None
/
–
The control box (BX3M-CB) is not connected.
Set the main switch to (OFF) and connect the control box and
the AC adapter of the control box again.
Then, set the main switch to (ON) and confirm that the beep is
heard once.
16, 100
, 117
None
/
–
The hand switch (BX3M-HS) is not connected.
Set the main switch to (OFF) and connect the hand switch
(BX3M-HS).
If the CUBE indicator or the OB indicator does not light even
though the hand switch is connected, contact Olympus.
16, 115
The hand switch (BX3M-HS) is damaged.
Contact Olympus.
–
None
/
–
–
The stored brightness (LIM function) is restored to the factory
default setting while the control box (BX3M-CB) is connected.
The power supply was turned ON while pressing the SET
switch of the microscope frame.
Set the main switch to (OFF) once and set it to (ON) again.
16
None
/
(Either one of them turns ON.)
–
The cable of the coded reflected light illuminator is not
connected.
Set the main switch to (OFF) and connect the cable of the
coded reflected light illuminator.
16, 115