76
: Turns ON. : Blinks. :Turns OFF. : Varies depending on state.
Phenomena
Cause
Remedy
Page
Number of beeps
CB
LIM /
HS
MIX
c) LIM is not working. (When following operations are performed, the stored brightness is not set automatically.)
· When selecting the objective by turning the motorized nosepiece or the coded nosepiece.
· When selecting the observation method using the coded reflected light illuminator.
None
–
/
–
–
The cables of the LED lamp housing, motorized nosepiece,
coded nosepiece or coded reflected light illuminator are
not connected.
Set the main switch to (OFF) and connect the cables of the
LED lamp housing, motorized nosepiece, coded nosepiece and
coded reflected light illuminator to the connector on the back of
the microscope frame.
16, 115
The brightness of LIM is not stored.
Operate according to the procedures.
27
The LED lamp housing is damaged.
If you cannot adjust the brightness even by pressing the LIM
switch on the front of the microscope frame to set to the Store
mode (The LIM indicator is OFF.), the LED lamp housing is
damaged. Contact Olympus.
27
None
–
/
–
–
The LIM is not set to the Replay mode.
Press the LIM switch on the front of the microscope frame to set
to the Replay mode. (The LIM indicator is ON.)
27
The microscope frame is damaged.
If the LIM indicator does not light by pressing the LIM switch
on the front of the microscope frame, the microscope frame is
damaged. Contact Olympus.
27
d) The motorized nosepiece does not rotate even though pressing the button of the hand switch for motorized nosepiece (BX3M-HSRE).
None
/
–
–
The control box (BX3M-CB) is not connected.
Set the main switch to (OFF) and connect the control box and
the AC adapter of the control box again.
Then, set the main switch to (ON) and confirm that the beep is
heard once.
16, 100
, 117
The control box (BX3M-CB) is damaged.
If the indicator on the back of the controller does not light even
though the control box and the AC adapter of the control box
are connected again, and also if the beep sound is not heard
when the main switch is set to (ON), contact Olympus.
16, 100
, 117
3 times
/
–
–
The motorized nosepiece is not connected.
Set the main switch to (OFF) and connect the motorized
nosepiece again.
16, 104
The motorized nosepiece or its cable is damaged.
Contact Olympus.
–
None
/
–
–
The hand switch for motorized nosepiece (BX3M-HSRE) is
not connected.
Set the main switch to (OFF) and connect the hand switch for
motorized nosepiece (BX3M-HSRE).
If the indicator4 on the back of the control box does not turn
ON even though the hand switch for motorized nosepiece is
connected, contact Olympus.
16, 115, 100
The hand switch for motorized nosepiece (BX3M-HSRE) is
damaged.
Contact Olympus.
–
None
/
–
–
The hand switch for motorized nosepiece (BX3M-HSRE) or
the motorized nosepiece is damaged.
Contact Olympus.
–
The stored brightness (LIM function) is restored to the factory
default setting while the control box (BX3M-CB) is connected.
The power supply was turned ON while pressing the SET
switch of the microscope frame.
Set the main switch to (OFF) once and set it to (ON) again.
16