72
Phenomena
Cause
Remedy
Page
j) The one-sided blur appears in
the observation image.
The nosepiece is not attached
correctly.
Push in the nosepiece along the
mounting dovetail until it touches the
end, and secure it.
104
The objective is not correctly
engaged in the light path.
Turn the nosepiece until the clicking
sound is heard and engage the
objective in the light path.
23
The sample is not parallel to the
stage.
Correct the sample to parallel to the
stage or change the sample.
19
The sample exceeds the specified
weight.
Replace with another sample.
19
k) The observation image shifts
when defocusing.
The nosepiece is not attached
correctly.
Push in the nosepiece along the
mounting dovetail until it touches the
end, and secure it.
104
The objective is not correctly
engaged in the light path.
Turn the nosepiece until the clicking
sound is heard and engage the
objective in the light path.
23
The aperture diaphragm is narrowed
down the centering is not performed
yet.
Perform the centering of the aperture
diaphragm.
39
The stage is not mounted correctly.
Mount the stage correctly.
97
l) During the fluorescence
observation, the background
becomes brighter.
This is caused by the in-house
fluorescence of the transmitted light
illumination optical systems.
Insert the light shield sheet in the
space of the stage.
18