a
Layout of the cable for coded reflected light illuminator
and the cable for coded nosepiece
Attach the cable holder at 1 position ( a ) on the right side of the back
of the microscope.
Open the cable holder a , insert the cable for coded reflected light
illuminator and the cable for coded nosepiece into the cable holder,
and close the cable holder.
1
2
116
Layout of the cable for U-MIXR
Attach the cable holder at 2 positions ( a , b ) on the right side and
1 position ( c ) on the back of the microscope respectively.
The cable holders are provided with the cable for U-MIXR
(U-MIXRCBL).
Open the cable holders a , b and c , put the cable for MIX slider
for reflected light observation (U-MIXR) into the cable holders, and
close the cable holders.
The cable holders (3 pcs.) are provided with the cable for U-MIXR
(U-MIXRCBL).
1
TIP
2
b
a
d
c
If you disconnect the connector from the MIX slider for
reflected light observation, store the cable in the cable holder
as shown in d .
When the connector is disconnected, the indicator e of the
MIX slider for reflected light observation turns OFF.
Indicator of MIX slider for reflected light observation
When the connector is connected
Turns ON
When the connector is not connected
Turns OFF
TIP
e