![Olympus BX-RLAA Instructions Manual Download Page 22](http://html1.mh-extra.com/html/olympus/bx-rlaa/bx-rlaa_instructions-manual_3904847022.webp)
19
U-DICRH
1. Adjust the background contrast by turning the prism movement knob |
on the DIC slider as described below. (Fig. 13)
2. When the prism movement knob on the U-DICRH DIC slider is turned, the
interference color in the background varies from -100 to 100 nm. Set the
retardation which can provide best contrast.
· Selecting a gray background color enables 3D-looking observation with
high contrast thanks to the highest sensibility of the gray sensitive color.
· Selecting a magenta sensitive color allows even small phase variation to
be observed as a change in color.
To select the magenta sensitive color as the background color, use the U-
POTP3 polarizer and insert it so that the indication can be seen from the
front.
# Since the DIC observation has a high detection sensitivity, take care
against contamination on the specimen surface.
}The detection sensitivity is variable depending on direction so it is recom-
mended to use a rotary stage.