Phase Contrast on the Olympus BH-2 Microscopes                                                  

   Revision 1                                                   

 

Page 5 of 6 

Center the Selected Phase Annulus 

Per  the  following  procedure,  carefully  adjust  the  two 
orthogonal,  spring-loaded  centering  thumbscrews  on 
the rear of the phase contrast condenser (see

 

Figure 6

) 

in order to center the bright image of the phase annulus 
within the dark image of the phase ring in the objective 
(se

Figure 7

 and 

Figure 8

). 

 

Figure 6 

 Centering screws of BH2-PC/PCD condenser 

 

Figure 7 

 The phase ring and phase annulus images 

To  adjust  the  position  of  the  phase  annulus,  carefully 
depress 

both 

of 

the 

spring-loaded 

centering 

thumbscrews  on  the  rear  of  the  phase  contrast 
condenser (se

Figure 6

until the thumbscrews engage 

with  the  internal  centering  screws  for  the  selected 

phase  annulus.    Then,  while  watching  through  the 
phase-centering  telescope,  adjust  the  thumbscrews  to 
get a feel for how the centering adjustments work and 
how  the  two  thumbscrews  interact  with  each  other.  
Once you have gained a feel for it, use the thumbscrews 
to adjust the position of the selected phase annulus so 
that its image (the bright ring) falls within the image of 
the phase ring in the objective (the dark ring) as shown 
in 

Figure 8

.   

 

Figure 8 

 The phase annulus properly centered 

After  the  selected  phase  annulus  has  been  properly 
centered,  release  the  two  centering  thumbscrews  and 
allow  them  to  disengage  from  the  internal  centering 
screws for the phase annulus. 

Caution

: Never turn the thumbscrews counterclockwise 

beyond  the  point  where  the  phase  annulus  ceases  to 
move in the field of view, or clockwise beyond the point 
where  their  rotation  begins  to  feel  firm,  and  do  not 
rotate the selector dial on the phase contrast condenser 
while one or both centering thumbscrews are engaged 
with  the  internal  centering  screws,  otherwise  damage 
to the phase contrast condenser may result

Center the Remaining Phase Annuli 

Repeat  the  centering  procedure  described  above  to 
center  the  phase  annuli  in  the  condenser  for  the 
remaining phase contrast objectives on the microscope.   

Replace Centering Telescope with the Eyepiece 

Once  the  various  phase  annuli  have  been  properly 
centered,  remove  the  phase-centering  telescope  from 
the  right-hand  ocular  tube  and  replace  it  with  the 
eyepiece. 

Place Green/Yellow Filter into Filter Receptacle 

Carefully  place  the  green/yellow  phase  contrast  filter 
(the IF-550 interference filter provides the best results, 
but  lacking  this,  the  45G533  absorptive  filter  may  be 
used  instead)  into  the  45mm  filter  receptacle  beneath 
the stage (see 

Figure 9

)