Phase Contrast on the Olympus BH-2 Microscopes                                                  

   Revision 1                                                   

 

Page 4 of 6 

 

Olympus IF-550 (interference) or 45G533 (absorptive) filter 

 

One or more Olympus LB phase contrast objectives 

Match Selector Dial with the Low-Power Objective 

Select  the  lowest-powered  phase  contrast  objective 
available on your  microscope.   Rotate  the  selector dial 
on  the  front  of the  phase  contrast  condenser  until  the 
number  visible  on  the  front  of  the  dial  matches  the 
selected phase contrast objective (se

Figure 1

).  Make 

sure the selected number on the dial clicks in place. 

 

Figure 1 

 Set condenser dial to match the objective 

Replace Eyepiece with Phase-Centering Telescope 

Carefully  remove  the  eyepiece  from  the  right-hand 
ocular  tube  and  insert  an  Olympus  phase-centering 
telescope

1

 in its place (see 

Figure 2

). 

 

Figure 2 

 Replace right eyepiece with phase telescope 

Focus the Phase-Centering Telescope 

Look  into  the  phase-centering  telescope.    With  the 
lighting intensity set to a comfortable viewing level, you 

 

1

  Use  the 

“CT

-

5”  for  23mm

  tubes  and  use  the 

“CT

-

30”  for  30mm 

tubes.  Note that other 

manufacturer’s centering telescope

s may be 

substituted for the Olympus CT, so long as the barrel diameter of the 
phase telescope is compatible with your ocular tubes. 

should see a  bright  ring of light  and  also  a darker ring, 
similar to that shown in 

Figure 3

.  Note that the image 

may be so blurry  that  these  features cannot be  readily 
discerned  at  this  point.    Also  note  that  the  relative 
positioning of the two rings may vary from that shown 
in 

Figure 3

.  

 

Figure 3 

 Phase telescope improperly focused 

Rotate the knurled focusing ring on the phase-centering 
telescope  (see 

Figure  4

)  until  the  image  seen  through 

the  phase-centering  telescope  is  sharply  focused  (see 

Figure 5

). 

 

Figure 4 

 Adjust knurled focus ring to focus the image 

 

Figure 5 

 The phase telescope is properly focused 