background image

Phase Contrast on the Olympus BH-2 Microscopes                                                  

   Revision 1                                                   

 

Page 6 of 6 

 

Figure 9 

 Place green/yellow filter into filter receptacle 

Congratulations,  you  have  now  properly  setup  your 
BH-2 microscope for phase contrast viewing. 

Using Phase Contrast 

The  hard  part  is  now  done.    From  here  on  out,  to  use 
phase  contrast,  just  be  sure  to  always  match  the 
number  on  the  selector  dial  on  the  front  of  the  phase 
contrast  condenser  with  the  magnification  of  the 
selected  phase  contrast  objective.    That  is  basically  all 
there is to it.  D

on’t forget to return the selector dial on 

the  phase  contrast 

condenser  to  the  “0” 

setting 

whenever you wish to switch back to regular brightfield 
viewing. 

Periodically Verify Centering of the Phase Annuli 

It  is  good  practice  to  re-check  the  centering  of  the 
phase  annuli  before  starting  a  phase  contrast 
observation  session.    With  practice,  you  will  come  to 
see  that  once  the  phase  annuli  have  been  centered, 
they rarely need to be re-adjusted. 

About the Phase Contrast Filter 

Note that it is necessary to use the green/yellow filter in 
the illumination system, in order to provide light of the 
proper  wavelengths  with  which  the  phase  contrast 
objectives  were  designed  to  operate.    This  is  critical  if 
you wish to obtain optimal phase contrast results.  This 
filter  is  particularly  important  when  you  wish  to 
perform  photomicrography, 

since  that’s  when  image 

quality is most critical.  The inevitable green/yellow cast 
that the filter imparts may be removed from the images 
by converting them to grayscale. 
For  routine  viewing,  however,  you  may  find  that  you 
prefer  to  omit  the  green  filter  from  the  illumination 
system, knowing that the phase contrast effects will be 
less than ideal, but the resulting color rendition will be 
much closer to normal. 

Phase Contrast Artifacts 

Certain  visual  artifacts  will  inevitably  appear  when 
viewing  specimens  under  phase  contrast,  which  can 
sometimes  make  it  difficult  to  accurately  interpret  the 
resulting images.  The bad news is you really cannot get 
rid  of  these  artifacts.  You  will  find  that  certain 
specimens  are  ideally  suited  for  phase  contrast, 
whereas  others  are  better  suited  to  regular  brightfield 
viewing, because of these artifacts.   
Halos: The most obvious phase contrast artifacts are the 
halos  which  you  will  see  surrounding  your  specimens.  
These  halos  are  caused  by  some  of  the  diffracted  light 
from  the  specimen  passing  through  the  phase  ring  in 
the objective.  Ideally, only the background illumination 
should pass through the phase ring, and only diffracted 
light  should  pass  through  the  areas  inside  and  outside 
the phase ring.  In positive phase contrast imaging (i.e., 
when  using  PL  or  PLL  phase  contrast  objectives),  this 
effect  causes  larger  objects  have  a  brighter  edge, 
whereas  in  negative  phase  contrast  (i.e.,  when  using 
NM or NH phase contrast objectives), this effect causes 
them to have a darker edge. 
Shade-Off  Effect:    Another  phase  contrast  artifact  is 
known  as  the 

Shade-Off  Effect

.    In  this  case,  the 

homogeneous parts of the image show up at the same 
brightness level as the background (i.e., the same as the 
surrounding  medium).    This  occurs  because  although 
the light experiences a phase shift as it passes through 
these regions of the specimen, only minimal diffraction 
occurs  and  the  angle  it  scatters  is  therefore  limited, 
causing it to pass through the phase ring and therefore 
not experience the desired interference. 
Contrast  Inversion:  A  third  phase  contrast  artifact  is 
known  as 

Contrast  Inversion

.    In  the  case  of  positive 

phase  contrast,  objects  with  a  high  index  of  refraction 
situated  next  to  objects  with  a  low  index  of  refraction 
will  appear  brighter  than  the  background,  instead  of 
darker.  This happens because in these cases the phase 
shift  is  not  the  usual  quarter  wave  that  should  occur, 
and  instead  of  the  expected  destructive  interference 
occurring, constructive interference occurs instead.  The 
opposite of this is true for negative phase contrast.  

How To Contact the Author 

Please  feel  free  to  direct  any  questions  or  comments 
regarding  this  document  (or  BH-2  microscopes  in 
general) to the author as listed on the cover page of this 
document. 
 
 

Reviews: