Test Programs
Maintenance Manual
3-10
Page
No. 5
Macro table
vertical
movement
check
<Macro table vertical movement check>
This program can be used to check the vertical movement of the Top Macro inspection table.
Press the [Start] button.
Select the operation speed using the menu buttons on the liquid crystal panel.
[M1]: High speed <Fast speed> [M2]: Medium speed <Middle speed> [M3]: Low speed <Slow
speed>
•
If you do not specify anything here, the loader will operate at the high speed [M1]: High
speed<Fast speed> setting.
Press the [Start] button and the loader starts the operation.
If the loader has not been initialized, initialization starts.
①
Raises the macro table
②
Lowers the macro table
The loader then repeats steps
①
to
②
.
•
Press the [Start] button to pause the operation at the break point.
•
The speed can be changed during the pause.
•
When an error occurs, the loader displays the same error code as in the regular operation, and then
stops.
•
Press the [Quit] button, and the macro table returns to its initial position and the loader stops the
operation.
Do not continue the operation for longer than 5 minutes.
Wait for longer than one minute before restarting the operation.
No. 6
L-arm
vertical
movement
check
<L-arm vertical movement check>
This program can be used to check the L-arm vertical movement.
Press the [Start] button.
Select the operation speed using the menu buttons on the liquid crystal panel.
[M1]: High speed <Fast speed> [M2]: Medium speed <Middle speed> [M3]: Low speed <Slow
speed>
•
If you do not specify anything here, the loader will operate at the high speed [M1]: High speed
<Fast speed> setting.
Press the [Start] button and the loader starts the operation.
If the loader has not been initialized, initialization starts.
①
Raises the L-arm (Lower limit
→
Middle point)
②
Raises the L-
arm (Middle point → Upper limit)
③
Lowers the L-
arm (Upper limit → Middle point)
④
Lowers the L-
arm (Middle point → Lower limit)
The loader then repeats steps
①
to
④
.
•
Press the [Start] button to pause the operation at the break point.
•
The speed can be changed during the pause.
•
When an error occurs, the loader displays the same error code as in the regular operation, and then
stops.
•
Press the [Quit] button, and the L-arm returns to its initial position and the loader stops the operation.
Do not continue the operation for longer than 5 minutes.
Wait for longer than one minute before restarting the operation.
CAUTION
CAUTION
Summary of Contents for AL120-12 Series
Page 2: ...This page intentionally left blank...
Page 13: ...Operation Maintenance Manual Introduction Conformity Standards i 11 Page...
Page 14: ...Introduction Conformity Standards Operation Maintenance Manual i 12 Page...
Page 16: ...Maintenance Manual Table of Contents C 2 Page This page intentionally left blank...
Page 59: ...Maintenance Manual Test Programs 3 31 Page This page intentionally left blank...