DMTA-10043-01EN, Rev. C, July 2016
Chapter 5
40
To perform the zero calibration
1.
Update the transducer zero compensation by wiping the transducer face clean of
all couplant and pressing
[2
nd
F]
,
[CAL ZERO] (Do
‑
ZERO)
while in
measurement mode.
2.
Couple the transducer to the standard.
3.
Press the
[CAL ZERO]
key.
4.
When the thickness reading is stable, press the
[ENTER]
key. The
[ENTER]
key
will not be accepted if the
LOS
flag is displayed.
5.
Uncouple the transducer and use the
[
]
,
[
]
,
[
]
, and
[
]
arrow keys to enter the
thickness of the standard.
6.
Press the
[MEAS]
key to complete the calibration and return to measurement
mode. If the gage is turned off before the
[MEAS]
key is pressed, the zero value
will not be updated to the new value, but instead will retain the previous value.
If the gage sounds a long beep before returning to measurement mode, an error has
been made in the calibration procedure and the zero value has not been changed. The
most likely cause is that the entered thickness was not correct.
Summary of Contents for 27MG
Page 6: ...DMTA 10043 01EN Rev C July 2016 Table of Contents vi...
Page 8: ...DMTA 10043 01EN Rev C July 2016 List of Abbreviations viii...
Page 22: ...DMTA 10043 01EN Rev C July 2016 Important Information Please Read Before Use 14...
Page 30: ...DMTA 10043 01EN Rev C July 2016 Chapter 1 22...
Page 38: ...DMTA 10043 01EN Rev C July 2016 Chapter 3 30...
Page 42: ...DMTA 10043 01EN Rev C July 2016 Chapter 4 34...
Page 50: ...DMTA 10043 01EN Rev C July 2016 Chapter 6 42...
Page 64: ...DMTA 10043 01EN Rev C July 2016 Chapter 7 56...
Page 76: ...DMTA 10043 01EN Rev C July 2016 Chapter 10 68...
Page 84: ...DMTA 10043 01EN Rev C July 2016 Chapter 11 76...
Page 88: ...DMTA 10043 01EN Rev C July 2016 Appendix 80...
Page 90: ...DMTA 10043 01EN Rev C July 2016 List of Figures 82...
Page 92: ...DMTA 10043 01EN Rev C July 2016 List of Tables 84...