DMTA-10043-01EN, Rev. C, July 2016
Measurements
41
6. Measurements
Once the initial setup of the 27MG Ultrasonic Thickness Gage has been performed
(see chapter “Initial Setup” on page 31) and a standard calibration has been
completed (see “Standard Calibration Measurement” on page 35), measurements can
be made.
To make measurements
1.
Apply couplant to the test block or material at the spot to be measured.
In general, the smoother the material surface, the thinner the couplant may be. Rough
surfaces require more viscous couplant such as gel or grease. Special couplants are
required for high temperature applications.
2.
Press the tip of the transducer to the surface of the material to be measured. Use
moderate to firm pressure and keep the transducer as flat as possible on the
material’s surface.
3.
Read the material’s thickness on the gage display.
For highest accuracy, both a velocity and zero calibration must be done.
Summary of Contents for 27MG
Page 6: ...DMTA 10043 01EN Rev C July 2016 Table of Contents vi...
Page 8: ...DMTA 10043 01EN Rev C July 2016 List of Abbreviations viii...
Page 22: ...DMTA 10043 01EN Rev C July 2016 Important Information Please Read Before Use 14...
Page 30: ...DMTA 10043 01EN Rev C July 2016 Chapter 1 22...
Page 38: ...DMTA 10043 01EN Rev C July 2016 Chapter 3 30...
Page 42: ...DMTA 10043 01EN Rev C July 2016 Chapter 4 34...
Page 50: ...DMTA 10043 01EN Rev C July 2016 Chapter 6 42...
Page 64: ...DMTA 10043 01EN Rev C July 2016 Chapter 7 56...
Page 76: ...DMTA 10043 01EN Rev C July 2016 Chapter 10 68...
Page 84: ...DMTA 10043 01EN Rev C July 2016 Chapter 11 76...
Page 88: ...DMTA 10043 01EN Rev C July 2016 Appendix 80...
Page 90: ...DMTA 10043 01EN Rev C July 2016 List of Figures 82...
Page 92: ...DMTA 10043 01EN Rev C July 2016 List of Tables 84...