23 

Guidance and Manufacture’s 

Declaration - Electromagnetic Immunity 

The Performance Table is intended for use in the electromagnetic environment specified below. The customer or the user 
should assure that it is used in such an environment. 

IMMUNITY Test 

IEC 60601 Test Level 

Compliance 

Level 

Electromagnetic Environment - Guidance 

 
Conducted RF 
IEC 61000-4-6 
 
 
 
Radiated RF 
IEC 61000-4-3 

 
3 V

rms

 

150 kHz to 80 MHz 
 
 
 
3 V/m 
80 MHz to 2,5 GHz 

 
3V

rms

 

 
 
 
 
3 V/m 

Portable and mobile RF communications equipment should 
not be used closer to any part of the Performance Table, 
including cables, than the recommended separation 
distance calculated by the equation applicable to the 
transmitter frequency. 
Recommended separation distance 

 
 
80 MHz to 800 MHz 
 
800 MHz to 2,5 GHz 
 
 

Where P is the transmitter maximum output power rating in 
watts (W) according to the transmitter manufacturer and d 
is the recommended separation distance in meters (m). 
Field strengths from fixed RF transmitters, as determined 
by an electromagnetic site survey 

a

, should be less than the 

compliance level in each frequency range. 

b

 

Interference may occur in the surroundings of equipment 
marked with the following symbol: 

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. 
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection 
from structures, objects, and people. 
 

 
 
 
 
 
 

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile 
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. 
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be 
considered. If the measure filed strength in the location in which the Performance Table is used exceeds the 
applicable RF COMPLIANCE LEVEL above, the Performance Table should be observed to verify normal operation. If 
abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the 
Performance Table. 
 
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. 

 

 
 

Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and the Performance 

Table 

The Performance Table is intended for use in an electromagnetic environment in which radiated RF disturbances are 
controlled. The customer or the user can help prevent electromagnetic interference by maintaining a minimum distance 
between portable and mobile RF communications equipment (transmitters) and the Performance Table as recommended 
below, according to the maximum output power of the communications equipment. 

Rated Maximum Output 

the Power of the 

Transmitter 

(W) 

Separation Distance According to Frequency of Transmitter (m) 

150 kHz to 80 MHz 

80 MHz to 800 MHz 

800 MHz to 2,5 GHz 

0,01 

0,12 

0,12 

0,23 

0,1 

0,37 

0,37 

0,74 

1,2 

1,2 

2,3 

10 

3,7 

3,7 

7,4 

100 

12 

12 

23 

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) 
can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power 
rating of the transmitter in watts (W) according to the transmitter manufacturer. 
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. 
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection 
from structures, objects, and people. 

Summary of Contents for PERFORMANCE TABLE

Page 1: ......

Page 2: ......

Page 3: ...Bracket Locks Optional 12 5 8 Double Bracket Locks 13 5 9 Articulating Armboards and Armrest for Blood Donation Optional 13 5 10 Knee Crutches Optional 14 5 11 Serum Support Optional 14 5 12 Rear Cas...

Page 4: ...0 10 Technical Features 21 10 1 Electromagnetic Compatibility EMC 22 11 Dimensional 24 12 Simbology 25 13 Important Notes 26 13 1 General Cares Compulsory Read 26 13 2 Discard 27 13 3 Transportation a...

Page 5: ...Auxiliary Pillow 1 Performance Table Quick Operation Guide 1 Interchangeable Armrests Opcionals Items 1 Multiarticulated Headrest 1 Wrist Rest Support 1 Articulating Armboards 1 Knee Crutches 1 Instru...

Page 6: ...ing Lights Optional Premium LED Operating Light 8 000 to 30 000 Lux Surgical LED Operating Light 30 000 Lux When the customer is purchasing the Performance Table he her can choose the articulating arm...

Page 7: ...uipped with up to 4 electric motors to perform up to 8 individual movements controlled by remote foot pedal or hand control It allows the user to program up to 8 work positions according to his needs...

Page 8: ...headrest 6 allows different adjustments for accommodation of the patient s head The wrist rest support 7 offers height adjustment allowing the professional to rest the wrists during procedures To repl...

Page 9: ...to pass through the metallic knob s axis 4 and then rotate it again 5 4 Performance Table Commands The Performance Table can be controlled via remote foot pedal 13 or hand control 12 Both can display...

Page 10: ...mmand of the other available numbers If the equipment has 4 motors the command 8 is preset at the factory setting the equipment on a stretcher position This command is not available for recording Note...

Page 11: ...for example press the command Seat Up when the seat is already in the highest position 5 4 7 Positioning of the Patient and the Operator During the equipment s movement the operator and other persons...

Page 12: ...height and tighten the plastic knob of the bracket 4 The instrument s tray also has free rotation on its axis and can be adjusted 5 7 Left and Right Bracket Locks Optional Used to attach accessories...

Page 13: ...on the side rails The plastic handles 1 allow height and rotation on the shaft distance the accessory to seat adjustment while the metal lever 3 allows the bracket lock adjustment on the side rails S...

Page 14: ...on the requested position 5 11 Serum Support Optional With the same concept of space utilization and diversification of the functions of the equipment it is possible to install the serum support 1 th...

Page 15: ...ttons 1 and 2 5 13 2 Venturi Saliva Ejector Developed for saliva ejection the Venturi ejector has cannula adapters 1 and 4 for fitting disposable cannulas or autoclavable metal cannulas It also has a...

Page 16: ...nomy approximately 3 hours of operation a period that may vary according to the amount of commands executed by equipment The LED 1 indicates the status of the battery according to the colors shown GRE...

Page 17: ...with soft foam and coated with flexible laminated PVC or leather which besides being resistant providing easy asepsis for the whole Mechanical structure manufactured in SAE 1020 rolled steel profiles...

Page 18: ...hould be closed daily at the end of the workday This valve can also be used to adjust the water pressure entering the equipment preventing that the water arrives with too much pressure on the water un...

Page 19: ...mooth arm movement 8 1 Equipment Positioning To position the equipment put it on the stretcher position backrest down and leg rest up then take it to the place where it will be used Perform the moveme...

Page 20: ...lean the spittoon bowl 1 remove the spittoon bowl s finishing 5 the strainer 4 and the detachable water spouts 2 and 3 for then remove the spittoon bowl For removal of the strainer 4 use a tweezer or...

Page 21: ...nuous operation Time On 30 s Time Off 5 min Premium LED operating light and surgical operating light continuous Premium LED operating light intensity 8 000 to 30 000 lux Surgical operating light inten...

Page 22: ...used in such an environment IMMUNITY Test IEC 60601 Test Level Compliance Level Electromagnetic Environment Guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV...

Page 23: ...ironment due to fixed RF transmitters an electromagnetic site survey should be considered If the measure filed strength in the location in which the Performance Table is used exceeds the applicable RF...

Page 24: ...24 11 Dimensional...

Page 25: ...25 12 Simbology According to IEC 60601 1 and IEC 60878 standards...

Page 26: ...the main water valve when equipped with water unit at the end of the working day It is recommended to replace the battery every 4 years to ensure maximum performance of the equipment This procedure mu...

Page 27: ...the best destination of each component without harm to the environment This equipment may have 2 lead acid batteries and they can not be disposed of with common waste because they present a risk of le...

Page 28: ...pen the main air valve 3 The equipment s drain hose is obstructed 3 Release bent wrinkled hose 4 The sewer is clogged 4 Provide sewage clearance 5 The hydro pneumatic system is locked 5 Contact Olsen...

Page 29: ...ccessories Use only Olsen original parts and accessories The use of non original components can compromise the performance of the equipment increasing its emissions or reducing its electromagnetic imm...

Page 30: ...ects f Connection to wrong voltage power supply 5 This warranty will cease when a In the normal course of its period of warranty b Make changes in the product not authorized by Olsen c Adulterations i...

Page 31: ...nd of low maintenance cost These characteristics have been achieved thanks to our competent and dedicated team which make me very proud for many reasons as they are always giving the best of their cre...

Page 32: ...____________________________________ __________________________________________ __________________________________________ __________________________________________ __________________________________...

Page 33: ...____________________________________ __________________________________________ __________________________________________ __________________________________________ __________________________________...

Page 34: ...____________________________________ __________________________________________ __________________________________________ __________________________________________ __________________________________...

Page 35: ...35...

Page 36: ...36 55 48 2106 6000 export3 olsen odo br Registration at Ministry of Health of Brazilian 10281300011 Techinician In Charge Eng Cleber da Costa CREA SC S1 116283 5 Cod 5400399 Rev 10 22 03 2021...

Page 37: ......

Page 38: ......

Page 39: ...cha Opcional 12 5 8 Ajustes de la Abrazadera Doble 13 5 9 Soporte Lateral para Brazos y Apoyabrazos para la Donaci n de Sangre Opcional 13 5 10 Pierneras Tapizadas Intercambiables Opcional 14 5 11 Sop...

Page 40: ...Caracter sticas T cnicas 21 10 1 Compatibilidad Electromagn tica EMC 22 11 Dimensional 24 12 Simbologia 25 13 Notas Importantes 26 13 1 Cuidados Generales 26 13 2 Descarte 27 13 3 Transporte y Almace...

Page 41: ...erformance 1 Almohada Auxiliar 1 Gu a R pida de Operaci n Mesa Performance 1 Apoyabrazos Tapizados Intercambiables Opcionales 1 Cabecera Multiarticulada 1 Aro de Apoyo 1 Soporte Lateral para Brazos 1...

Page 42: ...ras Opcional L mpara Premium LED 8 000 to 30 000 Lux L mpara Quir rgica LED 30 000 Lux En el momento de la compra de la Mesa Performance el cliente puede optar por el soporte lateral para brazos o pie...

Page 43: ...sta 8 movimientos individuales comandados por pedal o control de mano Permite que el usuario utilice hasta 8 posiciones de trabajo de acuerdo a sus necesidades adem s de la posici n cero autom tica la...

Page 44: ...ionar mejor la cabeza del paciente El aro de apoyo 7 ofrece ajuste de altura permitiendo al profesional reposar las mu ecas durante los procedimientos Para sustituir la cabecera actual por el aro de a...

Page 45: ...del eje de la perilla met lica 4 y luego girarla nuevamente 5 4 Comandos de la Mesa Performance La Mesa Performance puede ser comandada a trav s de pedal 13 o control de mano 12 Ambos pueden presenta...

Page 46: ...de solo 2 segundos entre los pasos 3 y 4 es decir cuando escuchar el primer beep largo paso 3 r pidamente presione cualquier comando de los n meros disponibles para grabar la posici n de trabajo Si n...

Page 47: ...do el asiento ya se encuentra en la posici n m s alta 5 4 7 Posici n del Paciente Operador y otras Personas Durante la realizaci n de los movimientos del equipo el operador y dem s personas cercanas d...

Page 48: ...te el perno del soporte para accesorios 4 La bandeja tambi n tiene giro libre sobre su eje pudiendo ser ajustada 5 7 Abrazaderas Izquierda y Derecha Opcional Utilizadas para la fijaci n de accesorios...

Page 49: ...rnos de ajuste de accesorios 1 permiten ajuste de altura y distancia del asiento mientras la varilla met lica 3 permite el ajuste de posici n en los carriles Simplemente gire en el sentido antihorario...

Page 50: ...oporte para Suero Opcional Con el mismo concepto de utilizaci n del espacio y la diversificaci n de las funciones del equipo es posible instalar el soporte para suero 1 a trav s de las abrazaderas del...

Page 51: ...aire y agua 1 y 2 5 13 2 Eyector Venturi Opcional Desarrollado para succi n de saliva el aspirador Venturi posee boquillas 1 y 4 para encaje de c nulas desechables o c nulas met licas autoclavables T...

Page 52: ...a de la red el ctrica En este caso el equipo emite el aviso Modo Bater a secci n 5 17 1 La autonom a es de aproximadamente 3 horas de operaci n per odo que puede variar de acuerdo con la cantidad de c...

Page 53: ...junto Estructura mec nica producida en estructura de acero SAE 1020 soldados por proceso MIG brinda al conjunto resistencia y durabilidad Pintura electrost tica se aplica a todas las partes met licas...

Page 54: ...actividades al final del d a La llave tambi n se puede utilizar para ajustar el vol men de agua para el llena vasos o taza de la escupidera La presi n de trabajo debe estar entre 2 8 y 4 0 bar con un...

Page 55: ...poseen fugas y tiene su accionamiento perfecto La l mpara se enciende y se apaga normalmente y presenta un movimiento suave del brazo 8 1 Posici n del Equipo Para posicionar el equipo col quelo en la...

Page 56: ...de la taza 1 para limpieza saque la terminaci n de la taza 5 la rejilla 4 y los conductos destacables 2 y 3 para entonces sacar la taza Utilice pinza o guantes para remoci n de la rejilla 4 para evita...

Page 57: ...o de operaci n Mesa Performance no continuo Tiempo On 30 s Tiempo Off 5 min L mpara Premium LED y Quir rgica continuo Intensidad de la l mpara Premium LED 8 000 a 30 000 lux Intensidad de la l mpara Q...

Page 58: ...ivas Descargas Electrost ticas IEC 61000 4 2 6 kV por contacto 8 kV por aire 6 kV por contacto 8 kV por aire Los pisos deber an ser de madera u hormig n o estar provistos de baldosas de cer mica Si el...

Page 59: ...cado por transmisores de RF fijos se deber a considerar la posibilidad de realizar una prueba electromagn tica in situ Si la intensidad de campo medida en el lugar donde habr de usarse la Mesa Perform...

Page 60: ...24 11 Dimensional...

Page 61: ...25 12 Simbologia La simbolog a est en conformidad con las normas IEC 60601 1 y IEC 60878...

Page 62: ...el disyuntor o desconecte el equipo de la red y cierre el registro de agua del consultorio cuando est equipado con unidad de agua al final del expediente Se recomienda sustituir la bater a cada 4 a o...

Page 63: ...l gicos debidamente identificados y de acuerdo con la legislaci n vigente Para el correcto desecho de este equipo y sus componentes y accesorios se recomienda que estos sean enviados a empresas especi...

Page 64: ...el sistema de alcantarillado 5 Bloqueo en el sistema hidroneum tico 5 Contacte el servicio t cnico autorizado Olsen En caso de duda o identificaci n de problema con el equipo que no est presentado en...

Page 65: ...Olsen Utilice solamente piezas originales Olsen El uso de piezas accesorios o componentes no originales puede comprometer el funcionamiento adecuado el equipo aumentando sus emisiones o reduciendo su...

Page 66: ...n a la red el ctrica de energ a con tensi n incorrecta 5 Esta garant a cesar a Por el decurso normal de su plazo de validez b Por alteraciones no autorizadas por Olsen en el producto c Por adulteraci...

Page 67: ...de mantenimiento muy bajo Estas cualidades se alcanzar n a trav s de un equipo competente y dedicado del cual me enorgullezco sobre todos los aspectos dispuesto a llevar a nuestros clientes lo mejor...

Page 68: ...____________________________________ __________________________________________ __________________________________________ __________________________________________ __________________________________...

Page 69: ...____________________________________ __________________________________________ __________________________________________ __________________________________________ __________________________________...

Page 70: ...____________________________________ __________________________________________ __________________________________________ __________________________________________ __________________________________...

Page 71: ...35...

Page 72: ...36 55 48 2106 6000 export3 olsen odo br Registro no Minist rio da Sa de 10281300011 Respons vel T cnico Eng Cleber da Costa CREA SC S1 116283 5 C d 5400399 Rev 10 22 03 2021...

Reviews: